PDF MANUAL

Manual for Atlas Copco Oxygen Generator OGP 5

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

Atlas Copco Oxygen Generator OGP 5

Atlas Copco Oxygen Generator OGP 5

Create an account to install this asset package.

Maintenance Plans for Atlas Copco Oxygen Generator Model OGP 5

Integrate maintenance plans directly into your work orders in MaintainX.

16000 Hourly/2 Yearly Oxygen Generator Maintenance

Check mark

Pre-Maintenance Safety Precautions

Check mark

Close the generator's outlet valve and the air inlet valve

Check mark

Ensure machine working pressure is 0 bar

Check mark

Switch off the generator

Check mark

Is the generator fully depressurized and purged with air?

Check mark

Vent oxygen to the outside atmosphere

Check mark

Ensure no smoking or open flames are in the vicinity of the oxygen generator

Check mark

Avoid closing valves too fast to prevent high gas velocities

Check mark

Inspection and Replacement Tasks

Run this procedureArrow right

5 Yearly Oxygen Sensor Replacement

Check mark

Pre-Maintenance Safety Precautions

Check mark

Close the generator's outlet valve and the air inlet valve

Check mark

Ensure machine working pressure is 0 bar

Check mark

Switch off the generator

Check mark

Is the generator fully depressurized and purged with air?

Check mark

Vent oxygen to the outside atmosphere

Check mark

Ensure no smoking or open flames are present near the oxygen generator

Check mark

Avoid closing valves too fast to prevent hazardous gas velocities

Check mark

Oxygen Sensor Replacement

Run this procedureArrow right

1 Yearly Oxygen Sensor Calibration

Check mark

Safety Precautions and Initial Setup

Check mark

Ensure the following safety precautions are followed

Check mark

Calibration Check

Check mark

Set the generator in Manual mode (Operation Mode)

Check mark

Close the ball valve (BA09) of the purity measurement (sample) line

Check mark

Apply calibration check gas at 600 ml/min and a pressure of 0.35 bar(g)

Check mark

Has the sensor reached a stable value for 5 minutes?

Check mark

Does the display value correspond to the calibration gas certificate?

Check mark

Disconnect the calibration check gas and reconnect the feedback line to the pressure regulator

Run this procedureArrow right

4000 Hourly/1 Yearly Oxygen Generator Maintenance

Check mark

Pre-Maintenance Safety Precautions

Check mark

Read and understand all safety precautions and recommendations.

Check mark

Close the generator's outlet valve and the air inlet valve.

Check mark

Ensure machine working pressure is 0 bar.

Check mark

Switch off the generator.

Check mark

Is the generator fully depressurized and purged with air?

Check mark

Vent oxygen to the outside atmosphere.

Check mark

Ensure no smoking or open flames are present near the generator.

Check mark

Maintenance Checks and Replacements

Run this procedureArrow right

1 Daily Controller Check

Check mark

Safety Precautions

Check mark

Close the generator's outlet valve and the air inlet valve

Check mark

Ensure machine working pressure is 0 bar

Check mark

Switch off the generator

Check mark

Is the generator fully depressurized?

Check mark

Purge thoroughly with air to remove all oxygen

Check mark

Vent oxygen to the outside atmosphere

Check mark

Ensure no smoking or open flames near the oxygen generator

Check mark

Avoid closing valves too fast to prevent high gas velocities

Run this procedureArrow right

Parts for Atlas Copco Oxygen Generator OGP 5

Access the parts list for your equipment in MaintainX.

Part Name
Part Number

Flow Sensor

CS VA420

Vessel

BF90PED11L

Flow Sensor

CS VA420

Vessel

BF90PED11L

Flow Sensor

CS VA420

Vessel

BF90PED11L

Unlock
efficiency with MaintainX Assist

MaintainX Assist is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.

Dollar Coin Icon

Lower Maintenance Costs

Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.

Thousands of companies manage their assets with MaintainX

DuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterrollDuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterroll

More from Atlas Copco

Explore Other Assets