MaintainX Asset Hub logo
EnterprisePricing
Tecan Austria

Tecan Austria Microplate Reader SPARK CYTO400

Need answers fast?
Explore the manual using AI.

Tecan Austria Microplate Reader SPARK CYTO400
Tecan Austria Microplate Reader SPARK CYTO400

Turn manuals into instant answers
with your AI-powered assistant
Turn manuals into instant answers
with your AI-powered assistant

PDF MANUAL

Manual for Tecan Austria Microplate Reader SPARK CYTO400

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

Tecan Austria Microplate Reader SPARK CYTO400

Tecan Austria Microplate Reader SPARK CYTO400

Create an account to install this asset package.

Maintenance Plans for Tecan Austria Microplate Reader Model SPARK CYTO400

Integrate maintenance plans directly into your work orders in MaintainX.

Microplate Reader Absorbance Module Testing

Check mark

Uniformity 96-Well Plate

Check mark

Ensure all materials are available

Check mark

Pipette 200μl of the reagent into the wells of a Greiner 96-well plate according to the plate layout

Check mark

Calculate the Uniformity (CV%) using the formula: Uniformity (CV%) = (SDOG * 100) / meanOG

Check mark

Is the calculated Uniformity (CV%) within acceptable limits?

Check mark

Quality Control of the NanoQuant Plate

Check mark

Ensure all materials are available

Check mark

Pipette 2 μl of the reagent onto all positions of the NanoQuant Plate

Check mark

Start the NanoQuant application and perform average blanking procedure over all wells (16 positions)

Run this procedureArrow right

Microplate Reader Injector Module Testing

Check mark

Material Preparation

Check mark

Ensure the following materials are available

Check mark

Injector Preparation

Check mark

Prime the injector with distilled water

Check mark

Weighing Procedure

Check mark

Weigh the empty plate and note the weight (mg)

Check mark

Inject 20 μl into 20 wells of the Greiner 96-well plate

Check mark

Weigh the plate immediately after injection and note the weight (mg)

Check mark

Environmental Conditions

Run this procedureArrow right

1 Daily Microplate Reader Injector Maintenance

Check mark

Inspection

Check mark

Inspect the syringes(s) for leaks

Check mark

Inspect the tubing for leaks

Check mark

System Flushing

Check mark

Flush the whole system with distilled or deionized water after each use

Check mark

Flush the whole system with distilled or deionized water when the syringe is not in use

Check mark

Syringe Maintenance

Check mark

Ensure syringes do not run dry for more than a few cycles

Run this procedureArrow right

Microplate Reader Cell Chip Adapter Maintenance

Check mark

Wear protective gloves, protective glasses, and protective clothing

Check mark

Empty the cell chip adapter and carefully remove the springs installed on the inside of the adapter lid

Check mark

Carefully wipe all outside surfaces of the adapter and the springs with a lint-free paper towel soaked with 70% ethanol

Check mark

Allow to dry

Check mark

Reinstall springs before using the adapter

Check mark

Do not use the cell chip adapter without springs! Measurement errors may result.

Run this procedureArrow right

1 Weekly Microplate Reader Injector Maintenance

Check mark

Preparation

Check mark

Flush the system

Check mark

Ethanol Rinsing

Check mark

Rinse the fully lowered syringes with 70% EtOH for 30 minutes

Check mark

Pump all the fluid from the syringe and tubing into a waste container

Check mark

System Rinsing

Check mark

Rinse the syringe/injector system with 70% EtOH

Check mark

Rinse the syringe/injector system with distilled or deionized water and leave the fluid pathway filled for storage

Check mark

Injector Needle Cleaning

Run this procedureArrow right

Unlock efficiency
with MaintainX CoPilot

MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.

Dollar Coin Icon

Lower Maintenance Costs

Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.

Thousands of companies manage their assets with MaintainX

DuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterrollDuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterroll

More from Tecan Austria

Explore Other Assets