MaintainX Asset Hub logo
EnterprisePricing
Schneider Electric

Schneider Electric Pressure Transmitter IDP10S-F

Need answers fast?
Explore the manual using AI.

Schneider Electric Pressure Transmitter IDP10S-F
Schneider Electric Pressure Transmitter IDP10S-F

Turn manuals into instant answers
with your AI-powered assistant
Turn manuals into instant answers
with your AI-powered assistant

PDF MANUAL

Manual for Schneider Electric Pressure Transmitter IDP10S-F

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

Schneider Electric Pressure Transmitter IDP10S-F

Schneider Electric Pressure Transmitter IDP10S-F

Create an account to install this asset package.

Maintenance Plans for Schneider Electric Pressure Transmitter Model IDP10S-F

Integrate maintenance plans directly into your work orders in MaintainX.

Pressure Transmitter Vent Plug Replacement

Check mark

WARNING: EXPLOSION HAZARD

Check mark

De-energize transmitters before removing threaded housing covers

Check mark

Ensure no flammable or combustible atmosphere is present before disconnecting equipment

Check mark

Remove the vent plug by unscrewing it counterclockwise

Check mark

Clean the angled sealing surface on the housing

Check mark

Lubricate the o-ring and the threads on the new plug with silicone lubricant

Check mark

Install the new plug into the housing by screwing it clockwise

Check mark

Torque the new plug to 0.9 ±0.02 N-m (8 ±0.2 lbf-in)

Check mark

Verify the torque applied to the new plug

Run this procedureArrow right

Pressure Transmitter SV Rerange Calibration

Check mark

Initial Setup

Check mark

Is the FoxCalTM feature enabled?

Check mark

Ensure the transmitter is stabilized at the final operating temperature.

Check mark

Calibration Process

Check mark

Perform two-point calibration if FoxCalTM is enabled

Check mark

Rezero the transmitter output to eliminate zero shifts from position effects.

Check mark

Check if underrange and overrange output values extend beyond 4 and 20 mA, respectively.

Check mark

Zero Reading Check (DP Transmitters Only)

Check mark

Return the output to linear mode for zero reading check.

Run this procedureArrow right

Pressure Transmitter Time in Service Meter Calibration

Check mark

Viewing Time in Service

Check mark

View the total days the transmitter has been in service using the 1 Device Setup > 2 Calibration > Time in Service Meter menu selection.

Check mark

View the user days since the last reset using the 1 Device Setup > 2 Calibration > Reset User Days menu selection.

Check mark

Resetting User Days

Check mark

Reset the user days to zero using the 1 Device Setup > 2 Calibration > Reset User Days menu selection.

Check mark

Calibration and Zeroing

Check mark

Perform a two-point calibration if FoxCalTM is enabled.

Check mark

Rezero the transmitter output once it has stabilized at the final operating temperature.

Check mark

Check underrange and overrange output values to ensure they extend beyond 4 and 20 mA, respectively.

Run this procedureArrow right

Pressure Transmitter One-Point Calibration

Check mark

Calibration Preparation

Check mark

Select calibration method

Check mark

Is the FoxCalTM feature enabled?

Check mark

Perform One-Point Calibration at LRV

Check mark

Perform one-point offset calibration at LRV

Check mark

Is the transmitter output stabilized at the final operating temperature?

Check mark

Rezero the transmitter output if necessary

Check mark

Post-Calibration Checks

Check mark

Are underrange and overrange output values extending beyond 4 and 20 mA?

Run this procedureArrow right

Pressure Transmitter Partial Proof Testing

Check mark

Preparation

Check mark

Ensure required tools are available

Check mark

Follow the required protocol to remove the device from online operation

Check mark

Testing the Milliamp Output

Check mark

Override the milliamp output to the High Failsafe setting

Check mark

Confirm the mA output corresponds to 21.0 mA

Check mark

Override the milliamp output to the Low Failsafe setting

Check mark

Confirm the mA output corresponds to 3.6 mA

Check mark

Post-Test Procedures

Run this procedureArrow right

Parts for Schneider Electric Pressure Transmitter IDP10S-F

Access the parts list for your equipment in MaintainX.

Part Name
Part Number

Heat Sink, 1 Inch

N1214YS

Heat Sink, 1 1/2 Inch

N1214YR

Calibration Screw Fitting

F0101ES

Silicone Lubricant

0048130

Grease, 1.75 Oz. Tube - Lubriplate For Transmitters With Aluminum Housing

X0118CC

Heat Sink, 1 Inch

N1214YS

Heat Sink, 1 1/2 Inch

N1214YR

Calibration Screw Fitting

F0101ES

Silicone Lubricant

0048130

Grease, 1.75 Oz. Tube - Lubriplate For Transmitters With Aluminum Housing

X0118CC

Heat Sink, 1 Inch

N1214YS

Heat Sink, 1 1/2 Inch

N1214YR

Calibration Screw Fitting

F0101ES

Silicone Lubricant

0048130

Grease, 1.75 Oz. Tube - Lubriplate For Transmitters With Aluminum Housing

X0118CC

Unlock efficiency
with MaintainX CoPilot

MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.

Dollar Coin Icon

Lower Maintenance Costs

Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.

Thousands of companies manage their assets with MaintainX

DuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterrollDuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterroll

More from Schneider Electric

Explore Other Assets