Thermo Fisher Ion Chamber RO-20
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Battery Replacement
Battery 1 Replacement
Check Battery 1 meter reading
Replace AA cells if necessary
Did you observe the polarity indicators when installing the AA batteries?
Battery 2 Replacement
Check Battery 2 meter reading
Replace 12V alkaline batteries if necessary
Did you observe the polarity indicators when installing the 12V batteries?
Long-term Inactivity
Ion Chamber Calibration
Preparation
Check the current air pressure
Is the air pressure suitable for calibration?
Calibration Procedure
Select the Zero position on the Function Switch and zero the meter
Remove the plastic cover on the side of the can
Position the RO-20 so that the ion chamber is in a gamma field of known intensity
Adjust the calibration control for the proper meter indication
Record the meter reading near the '4' mark for best accuracy
Dessicant Replacement
Desiccant Package Replacement
Check the color of the desiccant crystals
Slip the hose from the fitting on the old desiccant pack and onto the new one
Silica Gel Crystals Drying
Select drying method for silica gel crystals
Store dried crystals in tightly sealed containers
Desiccant Package Inspection
Ensure the snap-on cap hole is not covered or clogged
Chamber Assembly Maintenance
Circuit Check
Light Switch Check
Flip the light switch in both directions and confirm lamps are functional
Battery 1 Voltage Measurement
Measure the voltage of the AA cells with Function Switch set to Battery 1
Is the measured voltage within the expected range?
Battery 2 Voltage Measurement
Measure the voltage of the 3 ea 12V alkaline batteries with Function Switch set to Battery 2
Is the measured voltage within the expected range?
Access the parts list for your equipment in MaintainX.
1 Pcb Power Supply Unit Ro-20
SF-11546501
Range Knob
HDKN25
Zero Knob
HDKN4
Dual Mos Input Lmc662Ain
ICAOALMC662
Rubber Feet
MMRU7
1 Pcb Power Supply Unit Ro-20
SF-11546501
Range Knob
HDKN25
Zero Knob
HDKN4
Dual Mos Input Lmc662Ain
ICAOALMC662
Rubber Feet
MMRU7
1 Pcb Power Supply Unit Ro-20
SF-11546501
Range Knob
HDKN25
Zero Knob
HDKN4
Dual Mos Input Lmc662Ain
ICAOALMC662
Rubber Feet
MMRU7
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX