The Stryker Stretcher 2131 is a reliable and versatile medical transport solution designed for patient safety and comfort. This stretcher features advanced hydraulic systems and durable construction, making it ideal for emergency and hospital settings. Ensure optimal performance with regular maintenance and quality spare parts.
Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Stryker Stretcher 2131
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Base Angle Sensor Replacement - (Base)
Bed plugged into a wall outlet
Brakes set using one of the brake control locations or by using the manual brake pedal
Bed raised to the full up position and patient right siderails (foot end and head end) raised
Head end, enter, foot end, and right base cover removed
Two screws securing the base angle sensor to the base frame removed using a Phillips screwdriver
Properly grounded using an ESD system
Three clips holding the board cover unclipped
Cable unplugged from the board
Steps reversed to install new base angle sensor
Battery Replacement (Footboard)
Plug the bed into a wall outlet
Raise the bed to the full up position and level
Unplug the bed and turn the battery disconnect switch to the “OFF” (0) position
Using a small slotted screwdriver or a utility knife, remove the rear footboard label
Using a Phillips screwdriver remove the six screws from the top of the control panel which the label was covering
Raise the display up and using a stubby Phillips screwdriver, remove the three screws on the bottom
Remove the display housing and unplug the two cables from the function control board and set the display assembly on a bench to work on
Remove the battery and replace with new battery
Reverse the above steps to reassemble
Battery Replacement - (Litter)
Bed plugged into wall outlet
Brakes set using one of the brake control locations or by using the manual brake pedal
Bed raised to the full up position
Mattress assembly removed or folded back to expose the foot section
Two screws securing the electrical cover removed using a Phillips screwdriver
Bed unplugged from the wall outlet and battery switch turned “OFF” (0)
J12 connector unplugged from the CPU/Power board
Both batteries lifted up and stood upright
Wires unbolted from the battery posts using a 5/16” nutdriver and a 5/16” combination wrench
Bed Lift Actuator (Foot) Replacement - (Base)
Plug bed into wall outlet
Raise head end siderails to full up position and ensure brakes are activated
Remove head end base cover by pulling up and out
Remove center base cover by pulling up and out
Using diagonal pliers, cut the zip tie securing the actuator cable to the base frame
Using Needle-nose pliers, remove the rue clips from the clevis pins securing the actuator
Lower the bed all of the way down so the litter is supported by the base litter stops
Using the bed up/down controls, tap the up or down button to remove tension on the clevis pins and remove clevis pins
Unplug the cable quick connect and remove the actuator
Bed Lift Actuator (Head) Replacement - (Base)
Plug bed into wall outlet
Raise the head end siderails to the full up position and set the brakes using one of the brake control locations or by using the manual brake pedal
Remove head end base cover by pulling up and out
Remove center base cover by pulling up and out
Using diagonal pliers, cut the zip tie securing the actuator cable to the base frame
Using Needle-nose pliers, remove the rue clips from the clevis pins securing the actuator
If the bed will lower down, run it all the way down so the litter is supported by the base litter stop
If the bed will not lower all the way down; using a jack and a 2 x 4, take tension off of the actuator clevis pins by placing the jack and 2 x 4 between the bottom of the right side of the litter and the floor and jack up just enough to take the litter weight off of the base frame
Using the bed up/down controls, tap the up or down button to remove tension on the clevis pins and remove clevis pins
Access the parts list for your equipment in MaintainX.
100V Adapter
QDF27-1841
120V Auxiliary Outlet
QDF8024
12" Network Cable
QDF21-2895
12V 3 LEDs Night Light
QDF27-2025
200V Adaptor
QDF27-1842
100V Adapter
QDF27-1841
120V Auxiliary Outlet
QDF8024
12" Network Cable
QDF21-2895
12V 3 LEDs Night Light
QDF27-2025
200V Adaptor
QDF27-1842
100V Adapter
QDF27-1841
120V Auxiliary Outlet
QDF8024
12" Network Cable
QDF21-2895
12V 3 LEDs Night Light
QDF27-2025
200V Adaptor
QDF27-1842
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX