MaintainX Asset Hub logo
EnterprisePricing
Stryker

Stryker Stretcher 6500-201-000

Need answers fast?
Explore the manual using AI.

Stryker Stretcher 6500-201-000
Stryker Stretcher 6500-201-000

The Stryker Stretcher 6500-201-000 is a high-quality medical transport solution designed for patient safety and comfort. This stretcher features advanced hydraulic systems and durable construction, making it ideal for emergency and hospital settings. Regular maintenance ensures optimal performance and longevity of this essential equipment.

Turn manuals into instant answers
with your AI-powered assistant
Turn manuals into instant answers
with your AI-powered assistant

PDF MANUAL

Manual for Stryker Stretcher 6500-201-000

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

Stryker Stretcher 6500-201-000

Stryker Stretcher 6500-201-000

Create an account to install this asset package.

Maintenance Plans for Stryker Stretcher Model 6500-201-000

Integrate maintenance plans directly into your work orders in MaintainX.

12 Hourly / 6 Monthly Maintenance

Check mark

Inspect hoses and fittings for damage or wear

Check mark

Verify the hydraulic velocity fuse

Check mark

Measure and check load height

Check mark

Verify 'jog' function is operating

Check mark

Verify high speed retract is working

Check mark

Verify there is no damage or wear to either switch

Check mark

Verify both switches operate correctly

Check mark

Verify no bent, broken or damaged components

Check mark

Verify no damage or tears on cot grips

Run this procedureArrow right

24 Hourly / 12 Monthly Maintenance

Check mark

Ambulance cot and fastener fit and function properly

Check mark

Safety bar engages the vehicle safety hook properly

Check mark

Cylinder is adjusted so the lock nut is tight and the cot stops moving when it hits the dead stops

Check mark

Release handle returns to the stowed position

Check mark

All welds intact, not cracked or broken

Check mark

Warning labels present, legible

Check mark

Tires in good condition

Check mark

Optional wheel locks checked and adjusted as necessary

Check mark

Sign off on the maintenance check

Run this procedureArrow right

2 Hourly / Monthly Maintenance

Check mark

Verify the in-fastener shut-off is configured properly

Check mark

Check routing(s) and connection(s), verify there are no hanging wires

Check mark

Verify that the manual release functions properly

Check mark

Inspect the ambulance cot frame/litter

Check mark

Inspect patient restraints for proper function and no excessive wear

Check mark

Inspect the ambulance cot frame/base

Check mark

Verify all wheels secure, rolling and swiveling properly

Check mark

Verify the safety bar operates properly

Run this procedureArrow right

6 Hourly / 3 Monthly Maintenance

Check mark

All fasteners are secure

Check mark

No hydraulic fluid (red) leaks

Check mark

Motor mount fasteners are secure

Check mark

No hydraulic fluid leaks

Check mark

No leaks in the reservoir

Check mark

No damage or pinching of wiring harness, cables or lines

Check mark

No damaged connectors

Check mark

Base extends/retracts smoothly when the manual release is engaged

Check mark

With 100 lbs or more on the cot, the cot does not lower when the manual release is pulled

Run this procedureArrow right

Ambulance Cot Motion Maintenance

Check mark

All the restraint straps are securely buckled around the patient

Check mark

The cot can be in any position for rolling

Check mark

When rolling the cot with a patient on it, position an operator at the foot end and one at the head end of the cot at all times

Check mark

During transport, approach door sills and/or other low obstacles squarely and lift each set of wheels over the obstacle separately

Check mark

Sign off on the ambulance cot motion maintenance

Run this procedureArrow right

Parts for Stryker Stretcher 6500-201-000

Access the parts list for your equipment in MaintainX.

Part Name
Part Number

2" Adhesive Loop Pile

6082-001-085

Activation Switch

6500-001-130

Actuation Slide

6500-001-095

A Valve

6500-001-286

Backrest Pouch Option

6500-130-000

2" Adhesive Loop Pile

6082-001-085

Activation Switch

6500-001-130

Actuation Slide

6500-001-095

A Valve

6500-001-286

Backrest Pouch Option

6500-130-000

2" Adhesive Loop Pile

6082-001-085

Activation Switch

6500-001-130

Actuation Slide

6500-001-095

A Valve

6500-001-286

Backrest Pouch Option

6500-130-000

Unlock efficiency
with MaintainX CoPilot

MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.

Dollar Coin Icon

Lower Maintenance Costs

Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.

Thousands of companies manage their assets with MaintainX

DuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterrollDuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterroll

More from Stryker

Explore Other Assets