The Stryker Stretcher 6252 Stair-PRO is a high-performance medical transport solution designed for safe and efficient patient handling. This stretcher features advanced hydraulic systems and durable construction, ensuring reliability in emergency and non-emergency situations. Ideal for healthcare facilities, it enhances mobility and safety during patient transfers.
Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Stryker Stretcher 6252 Stair-PRO
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Checklist
All fasteners secure (reference all assembly drawings)
All welds intact, not cracked or broken
No bent or broken tubing or sheet metal
No debris in wheels
All wheels secure and rolling properly
Front casters secure, rolling and swiveling properly (if equipped)
Wheel locks hold wheels securely when on and clear the wheels when off
Chair unfolds and locks properly
No rips or cracks in seat or back rest
Limitations Cleaning
Warning: This procedure requires trained personnel with PPE!
Loosen the track belts and remove the rear wheels
Clean the track frame completely with rubbing alcohol
Use water at high pressure to rinse the belts
Allow the belts to completely dry before reassembly
Reassemble the track belts and the rear wheels
Test the performance of the chair using a simulated patient weight while descending a flight of stairs
If performance does not return to the original condition, what action is taken?
Sign off on the cleaning procedure
Monthly Inspection
Select the frequency of calls
Is chair in good condition?
Describe any issues found during inspection
Sign off on the chair inspection
Restraint Straps Maintenance
Chest Straps Configuration
Wrap each strap around the chair frame, insert the end through the loop on the end of the strap and pull it tight.
Pull the strap across the patient’s chest, lengthening the strap as necessary.
Buckle the strap.
Pull the loose end of the strap to tighten it securely around the patient.
Repeat for the second strap and the lap belt.
Ankle Strap Configuration
Remove the three plastic pieces from the strap.
Thread the strap through one of the plastic “D” rings and slide the ring down to the end of the strap up to the white label.
The Patient Down Stairs Maintenance
Roll the chair to the stairs and align it squarely with the edge of the first step.
Foot end operator - Extend the foot end lift handles by pushing the red release buttons and pulling the handles out until they stop. Release the buttons and verify the handles are locked.
Head end operator - Use one hand to pull the red upper control handle release cable while using the other hand to pull up and fully extend the handle. Release the cable and verify the handle is locked on both sides in the fully extended position.
Head end operator - Squeeze the red track release bar against the black cross tube. Relax your grip on the release bar and forcefully pull the Stair-TREADTM system to the fully extended position until both sides lock securely. Always verify both sides of the Stair-TREADTM system are locked by trying to fold it back up.
Operators face each other while descending the stairs.
Head end operator - Tilt the chair back just far enough to allow the Stair-TREADTM system to contact the floor.
Both operators - Maintaining the angle, guide the Stair-PRO® Stair Chair over the edge of the stairs, allowing the Stair-TREADTM system to engage the first step.
Both operators - Glide down the stairs until the treads are level across the edges of two or three steps.
Head end operator - Apply slight downward pressure on the extendable upper control handle while the foot end operator applies slight upward pressure on the foot end lift handles to keep the chair from rocking forward as it glides down the stairs.
Access the parts list for your equipment in MaintainX.
6" Molded Wheel
6060-002-045
Adjustable Caster Lock Pedal
6080-300-030
Back Rest, Molded ABS
6250-001-116
Bearing
0081-226-000
Bearing Spacer
6060-002-046
6" Molded Wheel
6060-002-045
Adjustable Caster Lock Pedal
6080-300-030
Back Rest, Molded ABS
6250-001-116
Bearing
0081-226-000
Bearing Spacer
6060-002-046
6" Molded Wheel
6060-002-045
Adjustable Caster Lock Pedal
6080-300-030
Back Rest, Molded ABS
6250-001-116
Bearing
0081-226-000
Bearing Spacer
6060-002-046
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX