Huntleigh Healthcare Pocket Doppler FD1
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Pocket Doppler Cleaning
Surface Cleaning
Wipe any fluids from the surface of the product using a clean dry cloth
Wipe with a cloth dampened in 70% Isopropyl Alcohol
Completely dry with a clean, dry lint free cloth
Contamination Handling
If the product has been contaminated, use the methods described for patient applied parts
Post-Cleaning Verification
Check that all external surfaces are clean and free of dirt and fluids
Pocket Doppler Check
Pre-Inspection Guidelines
Check with facility’s local infection control policy and medical equipment cleaning procedures.
Observe warnings and guidance on cleaning fluid labelling regarding use and PPE
Cleaning Instructions
Ensure detergent or disinfectant wipes are used with excess solution squeezed out.
Wipe off disinfectant using a cloth dampened with clean water.
Inspection of Components
Inspect the probe tip for cracks or damage.
Check the cable and connector for any cracking or intermittent behaviour.
Pocket Doppler Probes Cleaning
Low Risk Cleaning Procedure
Remove soiling from the probe
Wipe with a mild neutral detergent
Wipe with a cloth dampened in water
Completely dry with a clean lint free cloth
Medium Risk Cleaning Procedure
Follow Low Risk Cleaning Procedure
Wipe with a cloth dampened in Sodium Hypochlorite (1,000ppm)
Wait for two minutes
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX