Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Fresenius Medical Care North America Hemodialysis Machine 2008T BlueStar
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
4000 Hourly / 1 Yearly Hemodialysis Machine Maintenance
Perform the following Preventive Maintenance procedures every 12 months or 4000 hours of machine operation, whichever comes first.
Make copies of the Annual Preventive Maintenance Checklist provided at the back of this manual and use them to record the maintenance done.
Before starting the Annual Preventive Maintenance, select one of the following on Annual Preventive Maintenance Checklist:
3.1 FILTERS AND O-RINGS
Clean filters and replace O-rings as follows. Replace any filters that appear damaged or corroded.
Inlet Water Filter in the inlet connector of the water supply (See Figure 2, pg. 9).
Filters and O-rings in the concentrate (acid) and bicarbonate connectors (See Figure 4). Replace all three O-rings in each connector.
Clean the Dialysate Inline filter (See Figure 2, pg. 9).
Replace the O-rings in the Dialysate Line Connectors (See Figure 2, pg. 9).
6 Monthly Alarm Operation And Pressure Holding Tests
Warning! The use of a 'test drip chamber' or 'dummy drip chamber' must never be used on the treatment floor. It must only be used in a controlled technical environment.
Place the dialysate lines in the shunt and close the door
Place a venous chamber filled with water in the holder on the level detector module
Place the machine in Dialyze mode and start the blood pump. Clear all blood and water alarms
Select the Test & Options button. On this screen select the Both Tests button. Press the [CONFIRM] key to start
Observe the machine stepping through the following alarm tests and ensure that each alarm produces all of the responses shown in Table 1
When the tests shown in Table 1 are complete, the machine conducts a pressure holding test. When the test ends verify that the display screen reports TEST COMPLETE, indicating that all the tests were passed successfully. Press the [RESET] key
Pull the hydraulic out and remove the output tube from the UF pump. Install a 24' tube (part number 545325-10) to the empty output port. Route this tubing out the back of the machine so that it will not be kinked or pinched when the hydraulics is closed. Close the hydraulics and place the other end of tube into a collection container
Conduct steps 1-6 again
6 Monthly Filter Cleaning
2.1 FILTERS
Clean filters as follows. Replace any filters that appear damaged or corroded.
1. Inlet Water Filter in the inlet connector of the water supply (See Figure 2, pg. 9).
Warning! After cleaning or replacing the inlet filter screen, disinfect the water inlet line as described in the Operator's manual and in accordance with your Unit Policy.
2. Filters in the concentrate (acid) and bicarbonate connectors (See Figure 3).
3. Clean the Dialysate Inline filter (See Figure 2, pg. 9).;
6 Monthly Hemodialysis Machine Maintenance
Warning: This maintenance check requires trained personnel with PPE!
Any moisture detected on the floor of the hydraulic unit or any surfaces?
Any strain on the electrical cables?
Pre-UF Pump filter inspected for leaks or distortion?
UF pump Output check valves inspected for signs of wear, damage or leaking?
8-pin heater connections next to the distribution board checked and tightened?
Power plug inspected for loose or frayed wires?
Power cord inspected for nicks or cuts in the insulation?
Strain relief tightly secured to the power supply chassis?
8000 Hourly Deaeration Motor Brushes Replacement
Warning: This procedure requires trained personnel with PPE!
Deaeration motor/pump assembly removed from the machine
Upload a photo of the marked motor housing and motor cap
Two screws holding the motor cap to the motor housing removed
Motor cap removed to gain access to the motor brushes
Motor brushes replaced
Motor reassembled, aligning the marks
Two T-25 screws installed and tightened while holding the pieces together
Note: If the motor case and motor cap are not installed using the marks, the motor will run in reverse and flow errors will result.
Access the parts list for your equipment in MaintainX.
Bibag® Connector
640919
Bibag® Inlet Filter
M30225
Bicarbonate Connector (Blue)
579072
Bicarbonate Connector (Blue)
F40007101
Blood Pump Rotor Springs
650174
Bibag® Connector
640919
Bibag® Inlet Filter
M30225
Bicarbonate Connector (Blue)
579072
Bicarbonate Connector (Blue)
F40007101
Blood Pump Rotor Springs
650174
Bibag® Connector
640919
Bibag® Inlet Filter
M30225
Bicarbonate Connector (Blue)
579072
Bicarbonate Connector (Blue)
F40007101
Blood Pump Rotor Springs
650174
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX