Waters UV/Visible Detector 2489
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Flow Cell Removal
Precautions
Wear clean, chemical-resistant, powder-free gloves
Prevent scratching of the flow cell parts
Use a clean, lint-free cloth or similar clean surface in the work area
Replace the flow cell gaskets when cleaning, rebuilding, or replacing other flow cell components
Required Tools and Materials
Ensure availability of the following tools and materials
Flow Cell Removal Steps
Power-off the detector and disconnect the power cord
Flow Cell Maintenance
Preparation
Ensure eluents are well-degassed and flowing before powering on the detector.
Flushing Phase
Perform flushing of the flow cell.
Was flushing effective?
Removing and Cleaning Phase
Remove the flow cell for cleaning.
Clean the flow cell thoroughly.
Replace flow cell components as required
Lamp Replacement
Lamp Replacement Procedure
Check if the detector lamp fails to ignite repeatedly
Check if the detector fails to calibrate
Power-off the detector
Wait for 10 seconds
Power-on the detector
Invoke the lamp optimization software routine
Consult additional resources if needed
Lamp Removal
Safety Precautions
Allow the lamp to cool for 30 minutes before removing it
Keep the lamp in the housing when handling the lamp
Power-off the detector before changing the lamp
Wear eye protection that filters ultraviolet light
Lamp Extinguishing and Cooling
Extinguish the lamp using the keypad by pressing <Shift> <Lamp> and then pressing <Shift> <Lamp> again
Allow the lamp to cool for at least 30 minutes after extinguishing
Power Disconnection
Fuse Replacement
Safety Precautions
Power-off the detector
Unplug the instrument to avoid electric shock
Fuse Removal
Pinch the sides of the spring-loaded fuse holder
Withdraw the spring-loaded fuse holder with minimum pressure
Remove and discard both fuses
Fuse Replacement
Ensure new fuses are properly rated
Access the parts list for your equipment in MaintainX.
Ultra Performance LC/MS And HPLC/MS Systems
715001307
Ultra Performance LC/MS And HPLC/MS Systems
715001307
Ultra Performance LC/MS And HPLC/MS Systems
715001307
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX