The Honeywell Gas Detector MiniRAE 3000 is a portable, advanced gas detection device designed for industrial applications. This model offers reliable performance in detecting volatile organic compounds (VOCs) and ensures safety in hazardous environments through its robust features and easy-to-use interface.
Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Honeywell Honeywell Gas Detector MiniRAE 3000 MiniRAE 3000
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Alkaline Battery Adapter Replacement
Warning: To reduce the risk of ignition of hazardous atmospheres, recharge the battery only in areas known to be non-hazardous. Remove and replace the battery only in areas known to be non-hazardous.
Remove the three Philips-head screws to open the compartment.
Insert four fresh AA batteries as indicated by the polarity (+/-) markings.
Replace the cover. Replace the three screws.
Remove the Li-ion battery pack from the battery compartment by sliding the tab and tilting out the battery.
Replace it with the alkaline battery adapter
Slide the tab back into place to secure the battery adapter.
IMPORTANT! Alkaline batteries cannot be recharged. The instrument’s internal circuit detects alkaline batteries and will not allow recharging. If you place the instrument in its cradle, the alkaline battery will not be recharged. The internal charging circuit is designed to prevent damage to alkaline batteries and the charging circuit when alkaline batteries are installed inside the instrument.
Note: When replacing alkaline batteries, dispose of old ones properly.
Pump Sampling
Warning: Only trained personnel should perform this procedure
Enter the current energy consumption of the pump
Inlet connector is tight and the inlet tubing is in good condition
Enter the flow rate measured from the gas inlet probe
Flow rate is above 450 cc/min
Pump working properly?
If pump is not working properly, describe the issue
Sign off on the pump sampling
Lamp & Lamp Housing Maintenance
Is the lamp operational?
If the lamp is operational, proceed with cleaning. If not, proceed with replacement.
Clean the lamp window surface and the lamp housing with GC grade methanol using a cotton swab.
Check for any remaining film on the lamp window after cleaning.
Dry the lamp and the lamp housing thoroughly after cleaning.
CAUTION: Never touch the window surface with the fingers or anything else that may leave a film. Never use acetone or aqueous solutions.
Remove the lamp from the lamp housing if the lamp does not turn on.
Place the lamp O-ring onto the new lamp.
Insert the new lamp, avoiding contact with the flat window surface.
Battery Replacement
Turn off the instrument
Slide down the battery tab to unlock the battery
Remove the battery pack from the battery compartment by tilting it out
Replace a fully charged spare battery pack inside the battery compartment. Make sure the battery pack is oriented properly inside the compartment
Slide the capture tab back up to its locked position
PID Sensor Cleaning
Warning: This procedure requires trained personnel with PPE!
PID sensor module placed in GC grade methanol
Sensor cleaned in ultrasound bath for at least 15 minutes
Sensor dried thoroughly
Electrodes of the sensor not touched by hand
Lamp housing wiped off with methanol-soaked cotton swab
Sensor turned over for inspection
Sensor electrodes checked for corrosion, damage, or misalignment
Sensor electrode fingers adjusted if necessary
Access the parts list for your equipment in MaintainX.
Sensor Detector
023-3010-000
Alkaline Battery
MN1500
Alkaline Battery Adapter
059-3052-000
Lamp
050-0000-004
THP Sensor Module
023-3011-000-FRU
Sensor Detector
023-3010-000
Alkaline Battery
MN1500
Alkaline Battery Adapter
059-3052-000
Lamp
050-0000-004
THP Sensor Module
023-3011-000-FRU
Sensor Detector
023-3010-000
Alkaline Battery
MN1500
Alkaline Battery Adapter
059-3052-000
Lamp
050-0000-004
THP Sensor Module
023-3011-000-FRU
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX