MaintainX Asset Hub logo
EnterprisePricing
HAAG STREIT

HAAG STREIT Applanation Tonometer AT-900

Need answers fast?
Explore the manual using AI.

HAAG STREIT Applanation Tonometer AT-900
HAAG STREIT Applanation Tonometer AT-900

Turn manuals into instant answers
with your AI-powered assistant
Turn manuals into instant answers
with your AI-powered assistant

PDF MANUAL

Manual for HAAG STREIT Applanation Tonometer AT-900

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

HAAG STREIT Applanation Tonometer AT-900

HAAG STREIT Applanation Tonometer AT-900

Create an account to install this asset package.

Maintenance Plans for HAAG STREIT Applanation Tonometer Model AT-900

Integrate maintenance plans directly into your work orders in MaintainX.

Tonometer Check

Check mark

Warning: This check must be carried out once a month.

Check mark

Is the measuring prism properly positioned?

Check mark

Is the control weight correctly set?

Check mark

Did the repeated check yield correct results?

Check mark

Defective devices must be sent immediately to your Haag-Streit representative.

Check mark

Haag-Streit offers a repair and service package. For further information, please contact your Haag-Streit representative directly.

Check mark

Sign off on the Tonometer Check

Run this procedureArrow right

Drum Position Check

Check mark

Check the drum position 0

Check mark

Check position −0.05: The zero line on the measuring drum is moved downwards, opposite the index, by a single marking width. The feeler arm must move by itself all the way to the stop-piece, in the direction of the examiner.

Check mark

Check position +0.05: The zero line on the measuring drum is moved upwards, opposite the index. The feeler arm must move all the way to the stop-piece, on the patient's side.

Check mark

Check the drum position 2

Check mark

Check position 1.95: With a drum position of 1.95, the feeler arm must move from the free-movement area toward the limit stop in the direction of the examiner.

Check mark

Check position 2.05: With a drum position of 2.05, the feeler arm must move from the free-movement area toward the limit stop in the direction of the patient.

Check mark

Check the drum position 6

Check mark

Check position 5.9: The check point is at 5.9. Move the measuring drum's calibration mark 6 downwards, opposite the index, by a ½ interval. The feeler arm must move towards the examiner.

Check mark

Check position 6.1: The check point is at 6.1. Move the measuring drum's calibration mark 6 upwards, opposite the index, by a ½ interval. The feeler arm must move towards the patient.

Run this procedureArrow right

Applanation Tonometer Cleaning

Check mark

Warning: The preparation instructions provided do not apply to tonometer measuring prisms.

Check mark

Tonometer measuring prisms must be prepared in accordance with a different manual.

Check mark

Comply with the stipulated exposure time.

Check mark

Observe the manufacturer's safety instructions.

Check mark

Too strong or aggressive disinfectants or cleaning liquids e.g. hydrogen peroxide will damage the finish and coating of the device.

Check mark

Do not use sprays.

Check mark

Do not use any towels that drip.

Check mark

Wring out any soaked towels before use when necessary.

Check mark

Ensure that no liquid penetrates the device.

Run this procedureArrow right

Parts for HAAG STREIT Applanation Tonometer AT-900

Access the parts list for your equipment in MaintainX.

Part Name
Part Number

Reusable Tonometer Measuring Prism

7220316

Tonosafe Disposable Tonometer Measuring Prism

7220345

Control Weight

3300346

Reusable Tonometer Measuring Prism

7220316

Tonosafe Disposable Tonometer Measuring Prism

7220345

Control Weight

3300346

Reusable Tonometer Measuring Prism

7220316

Tonosafe Disposable Tonometer Measuring Prism

7220345

Control Weight

3300346

Unlock efficiency
with MaintainX CoPilot

MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.

Dollar Coin Icon

Lower Maintenance Costs

Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.

Thousands of companies manage their assets with MaintainX

DuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterrollDuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterroll

More from HAAG STREIT

Explore Other Assets