Stryker Pneumo Sure High Flow Insufflator
Stryker Pneumo Sure High Flow Insufflator

The Stryker Pneumo Sure High Flow Insufflator is a high-performance medical device designed for laparoscopic surgeries. This advanced insufflator ensures optimal gas flow and pressure control, enhancing surgical precision and patient safety. Ideal for healthcare facilities, it combines reliability with user-friendly features for efficient operation.

Turn manuals into instant answers
with your AI-powered assistant
Turn manuals into instant answers
with your AI-powered assistant

PDF MANUAL

Manual for Stryker Pneumo Sure High Flow Insufflator

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

Stryker Pneumo Sure High Flow Insufflator

Stryker Pneumo Sure High Flow Insufflator

Create an account to install this asset package.

Maintenance Plans for Stryker Pneumo Sure High Flow Insufflatornull

Integrate maintenance plans directly into your work orders in MaintainX.

1 Year Gas Flow Rate Test

Check mark

Warning: This test requires trained personnel with PPE!

Check mark

Test setup with open connection, without connected insufflation tube.

Check mark

Set a nominal gas flow rate of 15 l/min.

Check mark

Start insufflation: Press the START function field.

Check mark

Press the gas volume function field (o.0 | must be displayed).

Check mark

Start measuring for one minute.

Check mark

Stop the insufflation after one minute: Press the STOP function field.

Check mark

Enter the gas consumption

Check mark

Was the gas consumption at least 11-12 l?

Run this procedureArrow right

1 Year Insufflator Basic Function Test

Check mark

Remove insufflation tube from device

Check mark

Turn the device on using the On/Off switch

Check mark

Set to High Flow operating mode

Check mark

Check the displayed values

Check mark

Start insufflation by pressing the START function field

Check mark

Select the max. nominal gas flow

Check mark

Stop insufflation by pressing the STOP function field

Check mark

Press the gas volume function field

Check mark

Check if the basic function test of the device is complete

Run this procedureArrow right

1 Year Insufflator Safety Test

Check mark

Does the fuse correspond with the specifications indicated by the manufacturer?

Check mark

Are labels and stickers on device legible?

Check mark

Does the mechanical condition of the device allow for its safe use?

Check mark

Is the device clean to ensure proper and safe functionality?

Check mark

Measure leakage currents according to IEC 60601-1 / EN 60601-1

Check mark

Measure protective conductor resistance according to IEC 60601-1 / EN 60601-1. The protective conductor resistance is measured while device is connected to the power supply. The max. value is 0.2 S.

Check mark

Measure the insulation resistance with 500-700 V DC. The min. value is 50 MS

Check mark

The electric strength with high voltage cannot be measured.

Check mark

Sign off on the safety test

Run this procedureArrow right

1 Year Max. Device Pressure Test

Check mark

Select operating mode

Check mark

Select the max. nominal gas flow

Check mark

Connect a manometer and an open Veress cannula to the insufflation tube connection/adapter.

Check mark

Select the max. gas flow

Check mark

Start insufflation: Press the START function field. The manometer registers a pulsing pressure increase.

Check mark

Enter the maximum pressure registered by the manometer

Check mark

Stop insufflation: Press the STOP function field.

Check mark

Sign off on the pressure test

Run this procedureArrow right

1 Year Pressure Monitoring Test in High Flow Operating Mode

Check mark

Set to High Flow operating mode

Check mark

Set a nominal pressure of 15 mm Hg

Check mark

Set a nominal gas flow of 3 l/min

Check mark

Set a venting pressure of 3mm Hg in the Venting pressure limit configuration menu

Check mark

Use the syringe to generate a pressure of at least 19 mm Hg, which registers on the manometer

Check mark

Start insufflation by pressing the START function field. An acoustic warning sound is emitted with a pressure of more than 19 mm Hg (for 5 seconds) and the display depicts Overpressure

Check mark

Reduce the pressure. The warning ends when the pressure falls below 19 mm Hg (nominal pressure plus 4 mm Hg). Stop insufflation by pressing the STOP function field

Check mark

Use the syringe to generate a pressure of at least 30 mm Hg, which registers on the manometer

Check mark

Start insufflation by pressing the START function field. An acoustic warning sound is emitted without delay in case pressure exceeds 30 mm Hg and the display depicts Overpressure

Run this procedureArrow right

Parts for Stryker Pneumo Sure High Flow Insufflator

Access the parts list for your equipment in MaintainX.

Part Name
Part Number

CO2 Switching Valve for Insufflators

620030503

DISS Tube for House Gas Supply

620040003

Filter for Insufflation

Z0536-01

Fuse T

105208634

Gas Connection CO2 for House Gas Supply DISS

Z5020-01

CO2 Switching Valve for Insufflators

620030503

DISS Tube for House Gas Supply

620040003

Filter for Insufflation

Z0536-01

Fuse T

105208634

Gas Connection CO2 for House Gas Supply DISS

Z5020-01

CO2 Switching Valve for Insufflators

620030503

DISS Tube for House Gas Supply

620040003

Filter for Insufflation

Z0536-01

Fuse T

105208634

Gas Connection CO2 for House Gas Supply DISS

Z5020-01

Unlock efficiency
with MaintainX CoPilot

MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.