The Stryker Pneumo Sure High Flow Insufflator is a high-performance medical device designed for laparoscopic surgeries. This advanced insufflator ensures optimal gas flow and pressure control, enhancing surgical precision and patient safety. Ideal for healthcare facilities, it combines reliability with user-friendly features for efficient operation.
Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Stryker Pneumo Sure High Flow Insufflator
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
1 Year Gas Flow Rate Test
Warning: This test requires trained personnel with PPE!
Test setup with open connection, without connected insufflation tube.
Set a nominal gas flow rate of 15 l/min.
Start insufflation: Press the START function field.
Press the gas volume function field (o.0 | must be displayed).
Start measuring for one minute.
Stop the insufflation after one minute: Press the STOP function field.
Enter the gas consumption
Was the gas consumption at least 11-12 l?
1 Year Insufflator Basic Function Test
Remove insufflation tube from device
Turn the device on using the On/Off switch
Set to High Flow operating mode
Check the displayed values
Start insufflation by pressing the START function field
Select the max. nominal gas flow
Stop insufflation by pressing the STOP function field
Press the gas volume function field
Check if the basic function test of the device is complete
1 Year Insufflator Safety Test
Does the fuse correspond with the specifications indicated by the manufacturer?
Are labels and stickers on device legible?
Does the mechanical condition of the device allow for its safe use?
Is the device clean to ensure proper and safe functionality?
Measure leakage currents according to IEC 60601-1 / EN 60601-1
Measure protective conductor resistance according to IEC 60601-1 / EN 60601-1. The protective conductor resistance is measured while device is connected to the power supply. The max. value is 0.2 S.
Measure the insulation resistance with 500-700 V DC. The min. value is 50 MS
The electric strength with high voltage cannot be measured.
Sign off on the safety test
1 Year Max. Device Pressure Test
Select operating mode
Select the max. nominal gas flow
Connect a manometer and an open Veress cannula to the insufflation tube connection/adapter.
Select the max. gas flow
Start insufflation: Press the START function field. The manometer registers a pulsing pressure increase.
Enter the maximum pressure registered by the manometer
Stop insufflation: Press the STOP function field.
Sign off on the pressure test
1 Year Pressure Monitoring Test in High Flow Operating Mode
Set to High Flow operating mode
Set a nominal pressure of 15 mm Hg
Set a nominal gas flow of 3 l/min
Set a venting pressure of 3mm Hg in the Venting pressure limit configuration menu
Use the syringe to generate a pressure of at least 19 mm Hg, which registers on the manometer
Start insufflation by pressing the START function field. An acoustic warning sound is emitted with a pressure of more than 19 mm Hg (for 5 seconds) and the display depicts Overpressure
Reduce the pressure. The warning ends when the pressure falls below 19 mm Hg (nominal pressure plus 4 mm Hg). Stop insufflation by pressing the STOP function field
Use the syringe to generate a pressure of at least 30 mm Hg, which registers on the manometer
Start insufflation by pressing the START function field. An acoustic warning sound is emitted without delay in case pressure exceeds 30 mm Hg and the display depicts Overpressure
Access the parts list for your equipment in MaintainX.
CO2 Switching Valve for Insufflators
620030503
DISS Tube for House Gas Supply
620040003
Filter for Insufflation
Z0536-01
Fuse T
105208634
Gas Connection CO2 for House Gas Supply DISS
Z5020-01
CO2 Switching Valve for Insufflators
620030503
DISS Tube for House Gas Supply
620040003
Filter for Insufflation
Z0536-01
Fuse T
105208634
Gas Connection CO2 for House Gas Supply DISS
Z5020-01
CO2 Switching Valve for Insufflators
620030503
DISS Tube for House Gas Supply
620040003
Filter for Insufflation
Z0536-01
Fuse T
105208634
Gas Connection CO2 for House Gas Supply DISS
Z5020-01
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX