Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Eppendorf Freezer U101
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Freezer Cleaning
7.1 Cleaning
NOTICE! Risk of material damage
• Maintenance, adjustment and repair work should be carried out only by QUALIFIED, EXPERIENCED personnel who have been AUTHORIZED to undertake such work by Eppendorf or its authorized agents.
• Failure to use authorized service agents will invalidate the warranty.
7.1.1 Painted surfaces
All exterior paint work and inner doors should be cleaned using a solution of mild detergent in water. Do not use abrasive cleaners or solvents.
7.1.2 Interior and shelves
The interior panels and shelves are made of stainless steel. They may be cleaned using a recommended cleaning solvent, 70 % Isopropyl alcohol 30 % distilled water applied with a lint free cloth.;
12 Monthly Freezer Lubrication
NOTICE! Risk of material damage
Maintenance, adjustment and repair work should be carried out only by QUALIFIED, EXPERIENCED personnel who have been AUTHORIZED to undertake such work by Eppendorf or its authorized agents.
Failure to use authorized service agents will invalidate the warranty.
Are you a qualified, experienced personnel authorized by Eppendorf or its authorized agents?
Is the outer door hinges and the handle mechanism lightly lubricated using general-purpose oil or spray grease?
Sign off on the freezer lubrication
Electrical Components Maintenance
NOTICE! Risk of material damage
• Maintenance, adjustment and repair work should be carried out only by QUALIFIED, EXPERIENCED personnel who have been AUTHORIZED to undertake such work by Eppendorf or its authorized agents.
• Failure to use authorized service agents will invalidate the warranty.
7.2.5 Electrical components
WARNING! Risk of personal injury
• During routine maintenance, care must be taken to avoid any damage to the gaskets and sealing grommets of these enclosures; also check the gaskets and sealing grommets routinely to ensure their integrity.
Should any damage or deformity be detected, the gasket and/or sealing grommet must be replaced immediately.
• Failure to observe this safety warning will invalidate the warranty and could result in a dangerous situation.
7.2.5.1 Lamps
1 Monthly Door or Lid Seal Cleaning
NOTICE! Risk of material damage
Maintenance, adjustment and repair work should be carried out only by QUALIFIED, EXPERIENCED personnel who have been AUTHORIZED to undertake such work by Eppendorf or its authorized agents.
Failure to use authorized service agents will invalidate the warranty.
7.1.5 Door or lid seal
Be sure to treat the door or lid seal with care. Avoid damaging this seal in any way. The freezer cannot operate properly with a defective seal.
It is advisable to wipe both the seal and the surface against which it seals with a soft dry cloth once a month.
Seal and surface wiped with a soft dry cloth
Sign off on the door or lid seal cleaning
Inner Doors Replacement
NOTICE! Risk of material damage
• Maintenance, adjustment and repair work should be carried out only by QUALIFIED, EXPERIENCED personnel who have been AUTHORIZED to undertake such work by Eppendorf or its authorized agents.
• Failure to use authorized service agents will invalidate the warranty.
7.2.4 Replacing the inner door
1. Fully open the outer door of the freezer.
2. Fit door to hinge pins and close.
3. Check to ensure that inner door gasket is sealing against the freezer trim.
4. If required, adjust the latch retainer by loosening the screws and moving forward or backwards.
5. Close outer door.;
Access the parts list for your equipment in MaintainX.
U101 Upright Freezer, 115 V
U9420-0000
U101 Upright Freezer, 208 - 230 V
U9420-0002
U101 Upright Freezer, 230 V
U9420-0001
U535 Upright Freezer, 115 V
U9430-0000
U535 Upright Freezer, 208 - 230 V
U9430-0002
U101 Upright Freezer, 115 V
U9420-0000
U101 Upright Freezer, 208 - 230 V
U9420-0002
U101 Upright Freezer, 230 V
U9420-0001
U535 Upright Freezer, 115 V
U9430-0000
U535 Upright Freezer, 208 - 230 V
U9430-0002
U101 Upright Freezer, 115 V
U9420-0000
U101 Upright Freezer, 208 - 230 V
U9420-0002
U101 Upright Freezer, 230 V
U9420-0001
U535 Upright Freezer, 115 V
U9430-0000
U535 Upright Freezer, 208 - 230 V
U9430-0002
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX