Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Thermo Fisher Hot Plate HPA1910M-26
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Hot Plate Contact Points Cleaning
Safety Precautions
Refer servicing to qualified personnel.
Avoid contact with hot surfaces.
Ensure the hot plate is cool before starting maintenance.
Disconnect from the power supply prior to maintenance and servicing.
Cleaning Bimetal Contact Points
Disconnect hot plate from power supply.
Turn hot plate upside down.
Remove perforated cover by removing four screws.
Hot Plate Insulator Replacement
Warning!
Refer servicing to qualified personnel.
Caution: Hot Surface. Avoid Contact.
To Replace Insulator on Adjusting Shaft
Disconnect hot plate from power supply.
Turn hot plate upside down and remove perforated cover.
Loosen set screw on stop collar and screw adjusting shaft out to remove insulator.
Insert new insulator and screw the shaft in part way. (Recalibration is necessary—refer to recalibration instructions on page 12.)
Replace perforated cover.
Hot Plate Control Unit Replacement
Warning!
Control Unit Replacement Steps
Disconnect hot plate from power supply.
Turn hot plate upside down and remove perforated cover.
Remove stop collar by loosening set screw and remove the adjusting shaft from case.
Disconnect heater coil from control contact stud and disconnect white insulated wire from control unit.
Remove two screws holding control on top plate and remove control unit.
Install new control unit and secure with two screws.
Reinsert adjusting shaft and stop collar.
Hot Plate Calibration
Warning! Refer servicing to qualified personnel.
Caution: Hot Surface. Avoid Contact.
Disconnect hot plate from power supply.
Turn hot plate upside down and remove perforated cover.
Remove the knob from the adjusting shaft by removing set screw.
Turn the shaft until the contacts just close and slide knob over with the pointer in line with the FIRST mark on the dial plate.
Tighten the knob to the shaft with about 3/16" between skirt of the knob and the dial plate.
Turn the knob counterclockwise to the OFF mark on the dial and set the stop collar on the shaft with the set screw against the stop pin with about 1/4" between the collar and the control bracket.
Tighten set screw on stop collar and check for free rotation of the control shaft between stops.
Hot Plate Heater Coil Replacement
Warning! Refer servicing to qualified personnel.
Caution: Hot Surface. Avoid Contact.
Heater Coil Replacement Steps
Disconnect hot plate from power supply.
Turn hot plate upside down and remove perforated cover.
Remove nut and washer from contact stud, then remove one end of heater coil.
Remove remaining nut holding heater coil and remove heater coil.
Install new heater coil and secure with two nuts. Space approx. 1/8" from bimetal strip.
Replace perforated cover.
Access the parts list for your equipment in MaintainX.
Plate Aluminum Top
PT27X1
Element For HPA1910M-26 240V
EL27X2
Control
CN19X3
Shaft Control Adjustment
SF25X1
Cord Set For HPA1910M-26 240V
CR226X2
Plate Aluminum Top
PT27X1
Element For HPA1910M-26 240V
EL27X2
Control
CN19X3
Shaft Control Adjustment
SF25X1
Cord Set For HPA1910M-26 240V
CR226X2
Plate Aluminum Top
PT27X1
Element For HPA1910M-26 240V
EL27X2
Control
CN19X3
Shaft Control Adjustment
SF25X1
Cord Set For HPA1910M-26 240V
CR226X2
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX