MaintainX Asset Hub logo
EnterprisePricing
Bio-Rad

Bio-Rad Hemoglobin Testing System D10 DJ

Need answers fast?
Explore the manual using AI.

Bio-Rad Hemoglobin Testing System D10 DJ
Bio-Rad Hemoglobin Testing System D10 DJ

Turn manuals into instant answers
with your AI-powered assistant
Turn manuals into instant answers
with your AI-powered assistant

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

Bio-Rad Hemoglobin Testing System D10 DJ

Bio-Rad Hemoglobin Testing System D10 DJ

Create an account to install this asset package.

Maintenance Plans for Bio-Rad Hemoglobin Testing System Model D10 DJ

Integrate maintenance plans directly into your work orders in MaintainX.

Hemoglobin Testing System ATX Power Supply Replacement

Check mark

Remove the Electronics Module and Power Supply cover

Check mark

Release the Cable Anchor and remove the Power Supply output connector from the Power Board CN5

Check mark

Remove two mounting screws from each side of the supply

Check mark

Remove the supply and lay it upside down on the work bench

Check mark

Remove the four screws holding the mounting bracket to the bottom of the supply

Check mark

Install the mounting bracket on the new supply

Check mark

Ensure screws are not overly tightened

Check mark

Fasten the Power Supply to the Module with the four screws removed earlier

Check mark

Connect Power Supply Output Connector CN5 to the Power Board

Run this procedureArrow right

Power Surge Protection Board Replacement

Check mark

1) Perform steps 1 through 7 in 4.2.4 above:

Check mark

1. Remove the Printer from the plastic housing. There are no fasteners, the printer lifts up from the front.

Check mark

2. Disconnect Power and Communication cables from the rear of the printer and lift the Printer from the instrument.

Check mark

3. Remove the Top Cover. Loosen the two half turn fasteners. Slide the cover to the rear and lift to remove.

Check mark

4. Remove the Right side door from the instrument.

Check mark

5. Loosen the half turn fastener from the Buffer tray side cover. Lift the cover and remove it from the instrument.

Check mark

6. Loosen the two Knurled fasteners in the lower corners of the supply module cover. The one on the right can be accessed with a Philips screw driver as shown. The left fastener can be unscrewed by hand or right angle Philips driver. If the fastener is overly tight the Syringe module can be removed to gain access.

Check mark

7. The cover is attached at the top with two pins. Lift the cover up to clear the attachment pins and out of the instrument.

Check mark

8. Remove the three Power Supply Connectors then the four mounting screws, one in each corner. Remove the supply from the instrument.

Run this procedureArrow right

Touch Screen Board Replacement

Check mark

1. Open the Right and Left Side doors. Perform steps 2 and 3 to remove the front cover.

Check mark

2. Disconnect the two cables connected to the board.

Check mark

3. Remove the two 2mm hex head screws fastening the board to the module bracket.

Check mark

4. Installation is the reverse of the steps above.;

Run this procedureArrow right

Hemoglobin Testing System Motherboard Replacement

Check mark

Preparation

Check mark

Ensure you have a long handle 2.5 mm Hex Head Driver

Check mark

Place a piece of paper under the Electronic Module to catch any dropped screws

Check mark

Motherboard Removal

Check mark

Remove all cables from the CPU Motherboard

Check mark

Remove the 9 fastening screws securing the board to the Electronic Assembly using a 2.5 mm Hex Head Driver

Check mark

Remove the two Rack Loader port connector standoff posts from the rear of the instrument

Check mark

Slide the board towards the front of the instrument and remove it from the rear

Check mark

Motherboard Installation

Run this procedureArrow right

Hemoglobin Testing System Display Replacement

Check mark

1. Open the Right and Left Side doors. Perform steps 2 and 3 of Section 4.3.1 to remove the front cover.

Check mark

2. Release the cables from the anchor closest to the display. Disconnect the Communication cable from the Touch Screen Driver board.

Check mark

3. Remove the four screws holding the Display to the Display Chassis Bracket. Remove the cable from the Display Board. Remove the Display from the instrument.

Check mark

4. Place the Display face down on the work surface. The Touch Screen Flex Cable attached to the Display has a White Mylar side and a conductor side. The external cable connected to the Flex cable is not keyed. Mark the external cable to show which side is connected to the White side of the Flex Circuit cable. Disconnect the External Cable from the Flex Cable.

Check mark

5. Connect the External cable removed in step 4 to the Flex Cable on the replacement display. Connect so that the White side of the Flex Cable connection corresponds with the mark placed on the cable in step 4.

Check mark

6. Install the Display Board connector removed in step 3. Attach the display to the Module Bracket with the four mounting screws.

Check mark

7. Perform steps 1 and 2 in reverse to connect the display and install the Front and Side covers.;

Run this procedureArrow right

Parts for Bio-Rad Hemoglobin Testing System D10 DJ

Access the parts list for your equipment in MaintainX.

Part Name
Part Number

Kyospeed 2 Printer

2200664

Printer Mounting Tray

2200646

Sata Hard Drive

2200662

DVD RW Drive

2200663

USB Port

2200658

Kyospeed 2 Printer

2200664

Printer Mounting Tray

2200646

Sata Hard Drive

2200662

DVD RW Drive

2200663

USB Port

2200658

Kyospeed 2 Printer

2200664

Printer Mounting Tray

2200646

Sata Hard Drive

2200662

DVD RW Drive

2200663

USB Port

2200658

Unlock efficiency
with MaintainX CoPilot

MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.

Dollar Coin Icon

Lower Maintenance Costs

Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.

Thousands of companies manage their assets with MaintainX

DuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterrollDuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterroll

More from Bio-Rad

Explore Other Assets