Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

RAMVAC Dental Products Dental Vacuum System Bulldog QT
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
2000 Hourly Oil Replacement
Drain Used Oil
Place empty oil container (minimum 4 quarts) under oil drain tube.
Open oil drain valve.
Close oil drain valve after oil stops draining.
Is there water in the oil?
Contact your dealer or RAMVAC if water is present in the oil.
Check Oil Filter
Unscrew oil filter and remove filter element.
Is the oil filter element clean?
1 Daily Dental Vacuum System Cleaning
Preparation
Ensure the RAMVAC is still on before starting the cleaning process.
Rinsing Vacuum Lines
Rinse each high volume line with hot water (approximately one quart).
Rinse each saliva ejector line with hot water (a few ounces).
Applying Vacuum Line Cleaner
Aspirate a few ounces of dental vacuum line cleaner through each vacuum line
Select the cleaner used
Ensure the cleaner has the following qualities
1000 Hourly Oil Drip Rate Check
Preparation
Ensure the Bulldog vacuum unit is thoroughly warmed-up
Record the vacuum setting in inches of Hg
Record the ambient temperature in ºF
Drip Rate Check
Measure the drip rate at the Dripper Site Glass in drops per minute
Is the drip rate within the specified range?
Follow-up Actions
Contact dealer or RAMVAC if the drip rate is not as specified
2000 Hourly V-Belt Check
V-Belt Inspection
Inspect V-belt for wear
Is the V-belt cracked or frayed?
V-Belt Replacement
Replace V-belt if cracked or frayed
V-Belt Tension Adjustment
Does the V-belt squeak at start up?
Adjust V-belt tension if it squeaks at start up
Contact Information
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX