MaintainX Asset Hub logo
EnterprisePricing
Thermo Fisher

Thermo Fisher Radiation Monitor RM-20

Need answers fast?
Explore the manual using AI.

Thermo Fisher Radiation Monitor RM-20
Thermo Fisher Radiation Monitor RM-20

Turn manuals into instant answers
with your AI-powered assistant
Turn manuals into instant answers
with your AI-powered assistant

PDF MANUAL

Manual for Thermo Fisher Radiation Monitor RM-20

MAINTENANCE RESOURCES

Complete asset maintenance, one click away

Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.

Documents

Documents & Manuals

Find all the essential guides in one place.

  • CheckTensioning Guide

  • CheckBelt-diagram

  • CheckC-120 pulleys

  • + 13 more

Work Orders

Work Order Templates

Pre-built workflows to keep your asset running smoothly.

  • CheckDaily Electrical System Inspection

  • CheckReplace Roller and Pulley

  • CheckInstall Engine B-120

  • + 29 more

Procedures

Procedures

Integrate maintenance plans directly into your work orders.

  • CheckMotion Industries

  • CheckApplied Industrial Technologies

  • CheckElectrical Brothers

  • + 5 more

Parts

Parts

Access the parts list for your equipment in MaintainX.

  • CheckDrive Motor

  • CheckB2 Rollers

  • CheckTensioning System

  • + 40 more

Thermo Fisher Radiation Monitor RM-20

Thermo Fisher Radiation Monitor RM-20

Create an account to install this asset package.

Maintenance Plans for Thermo Fisher Radiation Monitor Model RM-20

Integrate maintenance plans directly into your work orders in MaintainX.

True Frequency Calibration

Check mark

Initial Setup

Check mark

Ensure necessary equipment is available

Check mark

Connect the pulser to the instrument through the capacitor

Check mark

RM-20 DISCR Control Setup

Check mark

Set the RM-20 DISCR control for 1 V between pins 10 and 11 of A103

Check mark

Set the ALARM SET control fully clockwise

Check mark

Set the GM/P switch to P

Check mark

Pulse Generator Connection and Calibration

Check mark

Connect the pulser to the DETECTOR connector on the front panel

Run this procedureArrow right

Detector Response Calibration

Check mark

Detector Setup

Check mark

Set the GM/P switch

Check mark

Adjust DISCR control to simulate increased gain, ensuring sensitivity is above internal noise level

Check mark

Voltage Adjustment

Check mark

Is the proper high voltage known?

Check mark

Set controls to the known proper voltage

Check mark

Set voltage low and connect the detector

Check mark

Plot voltage plateau curve

Check mark

Set high voltage to the best operating value

Run this procedureArrow right

High Voltage Indication Calibration

Check mark

Preparation

Check mark

Ensure the following equipment is available

Check mark

Connection Setup

Check mark

Connect the positive side of the high voltage voltmeter to the outboard end of C117, the high voltage filter capacitor.

Check mark

Connect the reference side to chassis ground at the lug on the DETECTOR connector.

Check mark

Set the range switch to the HV position.

Check mark

Calibration Adjustment

Check mark

Adjust the HV ADJ potentiometer R149 until the high voltage meter reads 2000 V.

Check mark

Adjust the HV CAL potentiometer R147 until the front panel meter indicates 2.0 kV.

Run this procedureArrow right

Radiation Monitor Disassembly

Check mark

Cover Removal

Check mark

Remove the 10-32 screw from each side of the unit

Check mark

Lift the cover off

Check mark

Circuit Board Removal

Check mark

Unplug connectors

Check mark

Remove knobs and nuts from board-mounted components

Check mark

Unscrew the AC input connector and line filter assembly from the rear panel

Check mark

Remove the two 4-40 screws at the rear corners of the board

Check mark

Is the battery installed?

Run this procedureArrow right

Parts for Thermo Fisher Radiation Monitor RM-20

Access the parts list for your equipment in MaintainX.

Part Name
Part Number

Power Cord 3-Wire, #18

WRAC4 (4)

Speaker

ADSP3 (6)

Lamp 14 V, 0.08 A

LPBU6 (5)

Fuse 3 Ag, 1/8 A, Slo Blo

FUSB2 (1)

Fuse 3 Ag, 1 A, Slo Blo

FUSB12 (1)

Power Cord 3-Wire, #18

WRAC4 (4)

Speaker

ADSP3 (6)

Lamp 14 V, 0.08 A

LPBU6 (5)

Fuse 3 Ag, 1/8 A, Slo Blo

FUSB2 (1)

Fuse 3 Ag, 1 A, Slo Blo

FUSB12 (1)

Power Cord 3-Wire, #18

WRAC4 (4)

Speaker

ADSP3 (6)

Lamp 14 V, 0.08 A

LPBU6 (5)

Fuse 3 Ag, 1/8 A, Slo Blo

FUSB2 (1)

Fuse 3 Ag, 1 A, Slo Blo

FUSB12 (1)

Unlock efficiency
with MaintainX CoPilot

MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.

Time Lapse Icon

Reduce Unplanned Downtime

Ensure your team follows consistent procedures to minimize equipment failures and costly delays.

Wrench Clock Icon

Maximize Asset Availability

Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.

Dollar Coin Icon

Lower Maintenance Costs

Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.

Thousands of companies manage their assets with MaintainX

DuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterrollDuracellAB InBevErgonTitan AmericaMeritUnivar SolutionsRivianCadrexMagnaMaasPresriteSoutheast PowerInterroll

More from Thermo Fisher

Explore Other Assets