Adam Equipment Analytical Balance PW 214
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Analytical Balance Maintenance
Safety Precautions
Use the AC adapter designed by the manufacturer for the balance.
Avoid overloading or dropping material onto the platform.
Do not spill liquids on the balance.
Static Electricity Management
Discharge the static electricity of the samples, if possible.
Wipe both sides of the pan and the top of the case with an anti-static agent.
Analytical Balance Manual Calibration
Calibration using Internal Calibration mass
Check the display is at zero. Tare if necessary
Calibration will begin automatically. When calibration is complete the balance will return to weighing
Calibration using External Calibration mass
Press [Esc] to abort the calibration at any time
Press the [Cal] key
Ensure the pan is empty and press [Enter] when 'LOAD 0' is displayed
Enter the value of the calibration mass (e.g., 100,000 g)
Place the selected mass on the balance and press [Enter]
Analytical Balance Automatic Calibration
Automatic Calibration Conditions
Has the internal temperature changed more than 0.5°C?
Has it been more than 4 hours since the last calibration or 15 minutes after power was applied?
Calibration Process
Is the balance at a stable zero?
Is the 'CAL' symbol flashing on the display?
Wait for the 5 second countdown to complete for calibration to start.
Press [Esc] to delay calibration by one minute if needed.
External Calibration
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX