Turn manuals into instant answers
with your AI-powered assistantTurn manuals into instant answers
with your AI-powered assistant
Get instant access to all the maintenance information you need. Empower technicians to perform preventive maintenance with asset packages, ready to use right out of the box.
Find all the essential guides in one place.
Tensioning Guide
Belt-diagram
C-120 pulleys
+ 13 more
Pre-built workflows to keep your asset running smoothly.
Daily Electrical System Inspection
Replace Roller and Pulley
Install Engine B-120
+ 29 more
Integrate maintenance plans directly into your work orders.
Motion Industries
Applied Industrial Technologies
Electrical Brothers
+ 5 more
Access the parts list for your equipment in MaintainX.
Drive Motor
B2 Rollers
Tensioning System
+ 40 more

Thermo Fisher Circulating Water Bath 2867
Create an account to install this asset package.
Integrate maintenance plans directly into your work orders in MaintainX.
Circulating Water Bath Cleaning
General Cleaning
Clean the bath with mild soapy water.
Rinse the bath with distilled water.
Add formaldehyde or quaternary ammonium germicide if algae or microorganisms are present.
Chemical Usage Precautions
Ensure the following chemicals are never used
Limit the use of necessary chemicals to a maximum of four hours.
Removing Discoloration
Wear heavy gloves and goggles for protection.
Temperature Calibration
Preparation
Ensure a new temperature probe is installed into the bath
Connect the temperature probe connector to J302 on CPU board
Is the power switch in the OFF position?
Entering Calibration Mode
Press and hold the SETPOINT key while turning Power ON
End the LED display test by pressing SETPOINT
Adjust display to read 0012 and press SETPOINT
Calibration Process
Temperature Sensor Replacement
Preparation
Drain water from bath
Remove diffuser pan from bath
Sensor Removal
Remove connector J302 from CPU board
Remove temperature sensor fitting from inside bath pan
Remove rubber grommet from temperature sensor
Carefully slide sensor through the control housing and remove
Installation and Calibration
Circulating Pump Replacement
Preparation
Drain water from bath
Pump Removal
Remove the two pump leads attached to power switch
Remove green wire from pump to ground lug
Loosen the two visible hose clamps with a screwdriver
Remove tubing from pump
Disconnect pump from pump bracket by removing two screws
Remove pump from bath
Hi-Limit Thermostat Replacement
Thermostat Removal
Remove four leads from thermostat terminals
Remove two screws from thermostat bracket and remove bracket
Accessing the Thermostat
Carefully tilt bath on its side
Remove eight screws from bottom plate and remove plate
Remove insulation from bottom of pan and between pan and control panel wall
Thermostat Extraction
Loosen two nuts from thermostat bulb bracket
Access the parts list for your equipment in MaintainX.
Quaternary Ammonium Germicide
3166250
Power Supply Pcb Assembly
3176677
Quaternary Ammonium Germicide
3166250
Power Supply Pcb Assembly
3176677
Quaternary Ammonium Germicide
3166250
Power Supply Pcb Assembly
3176677
MaintainX CoPilot is your expert colleague, on call 24/7, helping your team find the answers they need to keep equipment running.
Reduce Unplanned Downtime
Ensure your team follows consistent procedures to minimize equipment failures and costly delays.
Maximize Asset Availability
Keep your assets running longer and more reliably, with standardized maintenance workflows from OEM manuals.
Lower Maintenance Costs
Turn any technician into an expert to streamline operations, maintain more assets, and reduce overall costs.
Thousands of companies manage their assets with MaintainX