Consarc Dip TC Testing Procedure
This procedure verifies the accuracy of the dip thermocouple, ensuring precise temperature control for the vacuum furnace to guarantee proper melting and consistent metallurgical quality.
Consarc Dip TC Testing Procedure
This procedure verifies the accuracy of the dip thermocouple, ensuring precise temperature control for the vacuum furnace to guarantee proper melting and consistent metallurgical quality.
Required Materials
Push button as show in picture to start
Enter mV reading on multimeter for 3050 F
if the answer is "> 12.5 "
Select the FAIL option and notify maintenance supervisor
if the answer is "< 11.7 "
Select Fail and Notify Maintenance Manager
Enter mV reading on multimeter for 2700 F
if the answer is "> 10 "
Select FAIL and Notify Maintenance Supervisor
if the answer is "< 9 "
Select FAIL and Notify Maintenance Supervisor
Enter mV reading on multimeter for 2450 F
if the answer is "> 8.5 "
Select FAIL and notify maintenance supervisor
if the answer is "< 7.5 "
Select FAIL and notify maintenance supervisor
Source: Oerlikon Plymouth (Community Member)