OSP300 Relocation Detection Unit Battery Change
Use this procedure to replace the relocation detection unit batteries while keeping at least one battery connector attached at all times.
OSP300 Relocation Detection Unit Battery Change
This procedure is used to replace the relocation detection unit batteries in the OSP300 control box. It covers powering off, swapping the battery holders one connector at a time so the unit never loses power, and avoiding an accidental relocation detection alarm.
Procedure Max Score: 7
Two battery cable connectors are provided for the relocation detection unit battery holder. Change batteries while either one of the battery cable connectors are connected the relocation detection unit.
How to change relocation detection unit batteries
If the batteries are changed with all of the two battery cable connectors disconnected simultaneously while the main breaker is turned off, the relocation detection function is activated when the power is turned on again after battery change. Entering password is needed to cancel the relocation detection status. Thus, make sure to change batteries while either one of the battery cable connectors are connected to the unit. However, if the main breaker is turned on, the relocation detection function will not
Source: MPM, Inc. (Community Member)