CYE12-15 Monthly Calibration

This procedure outlines the monthly calibration process for the CYE12-15 machine. Using a certified voltage meter, the technician verifies that each power supply is delivering the correct output within specified tolerance ranges. Completing this calibration ensures accurate electrical performance, supports safe and reliable machine operation, and provides proper documentation for maintenance records.
TAC Manufacturing
12/10/2025

CYE12-15 Monthly Calibration

    Calibration Procedure

    40. Measure the voltage at 15.00vdc power supply. (14.8vdc - 15.2vdc)

    • CYE12-1540.PNG
    Meter name
    Unit

    50. Measure the voltage at 15.00vdc power supply. (14.8vdc - 15.2vdc)

    • CYE12-1550.PNG
    Meter name
    Unit

    60. Measure the voltage at 24.00vdc power supply. (23.8vdc - 24.2vdc)

    • CYE12-1560.PNG
    Meter name
    Unit

    70. Measure the voltage at 12.00vdc power supply. (11.8vdc - 12.2vdc)

    • CYE12-1570.PNG
    Meter name
    Unit

    80. Measure the voltage at 14.00vdc power supply. (13.8vdc - 14.2vdc)

    • CYE12-1580.PNG
    Meter name
    Unit

    120. Enter the ID# of the voltage meter that was used.

    130. Sign Off

    Click here to sign

Source: TAC Manufacturing (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.