AT24-06 Monthly Calibration
This procedure confirms that machine power supplies meet specified voltage tolerances. Using a calibrated meter, readings are measured and recorded to ensure equipment operates correctly and safely.
AT24-06 Monthly Calibration
Calibration Procedure
40. Measure the voltage at the 24.00vdc power supply. (23.8vdc - 24.2vdc)
Meter name
Unit
50. Measure the voltage at the 12.00vdc power supply. (11.8vdc - 12.2vdc)
Meter name
Unit
60. Measure the voltage at the 5.00vdc power supply. (4.8vdc - 5.4vdc)
Meter name
Unit
70. Inside the enclosure measure the voltage between the GND connection and positive 12vdc power connection. (11.7vdc - 12.2vdc)
Meter name
Unit
80. Measure the voltage between the GND connection and positive +5vdc power connection. (4.8vdc - 5.3vdc)
Meter name
Unit
90. Measure the voltage between the GND connection and positive 6.2vdc power connection. (5.8vdc - 6.5vdc)
Meter name
Unit
140. Enter the ID# of the voltage meter that was used.
150. Sign Off