PR 11.3 Calibration Validation Form 12"
This procedure is used to validate a 12-inch caliper at set measurement points so its accuracy is confirmed against traceable standards, and any caliper that fails is taken out of service.
PR 11.3 Calibration Validation Form 12"
This procedure outlines the validation of a 12-inch caliper against NIST-traceable gauge blocks at 68°F (20°C), in line with ASME B89.1.14-2018 and ISO 9001:2015 (7.1.5). It guides the technician through recording the instrument details and as-received condition, measuring four outside and three inside points, recording the error at each point and the as-left condition, and removing the caliper from service if any point fails.
INSTRUMENT INFORMATION
INSTRUMENT ID
MANUFACTURER
MODEL
SERIAL NUMBER
CALIPER TYPE
ENVIRONMENT & AS RECEIVED CONDITION
AS RECEIVED CONDITION
AS RECEIVED NOTES (REQUIRED IF FAIL)
AS LEFT CONDITION
OUTSIDE JAW (EXTERNAL) MEASUREMENTS
Outside Jaw Check: Measure NIST-traceable gauge blocks (or certified length standards) using the outside jaws. Complete 5 measurement points across the caliper’s range. The form automatically calculates Error, selects MPE, and returns PASS/FAIL.
Gauge Block Actual (From Lab) - Point 1
Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions
OUTSIDE MEASUREMENT #1 DETAILS
Measured Value - Point 1
Calculate Measured Error
Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)
Measurement Point 1 - PASS/FAIL
Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.001 in
if the answer is "oneOf [FAIL]"
OUTSIDE MEASUREMENT #2 DETAILS
Gauge Block Actual (From Lab) - Point 2
Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions
Measured Value - Point 2
Calculate Measured Error
Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)
Measurement Point 2 - PASS/FAIL
Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.001 in
if the answer is "oneOf [FAIL]"
OUTSIDE MEASUREMENT #3 DETAILS
Gauge Block Actual (From Lab) - Point 3
Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions
Measured Value - Point 3
Calculate Measured Error
Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)
Measurement Point 3 - PASS/FAIL
Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.001 in
if the answer is "oneOf [FAIL]"
OUTSIDE MEASUREMENT #4 DETAILS
Gauge Block Actual (From Lab) - Point 4
To Acquire 11" for the Validation Target, use the 10" Gauge Block (TSG-GB003) paired with the 1" Gauge Block (TSG-GB002)
Measured Value - Point 4
Calculate Measured Error
Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)
Measurement Point 4 - PASS/FAIL
Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.0015 in
if the answer is "oneOf [FAIL]"
INTERNAL MEASUREMENTS
INSIDE MEASUREMENT #1 DETAILS
Gauge Block Actual (From Lab) - ID Point 1
Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions
Measured Value - ID Point 1
Calculate Measured Error
Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)
Measurement ID Point 1 - PASS/FAIL
Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): •Tolerance: ±0.001 in
if the answer is "oneOf [FAIL]"
INSIDE MEASUREMENT #2 DETAILS
Gauge Block Actual (From Lab) - ID Point 2
Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions
Measured Value - ID Point 2
Calculate Measured Error
Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)
Measurement ID Point 2 - PASS/FAIL
Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): •Tolerance: ±0.002 in
if the answer is "oneOf [FAIL]"
INSIDE MEASUREMENT #3 DETAILS
Gauge Block Actual (From Lab) - ID Point 3
Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions
Measured Value - ID Point 3
Calculate Measured Error
Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)
Measurement ID Point 3- PASS/FAIL
Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): •Tolerance: ±0.002 in
if the answer is "oneOf [FAIL]"
FINAL DETAILS
DID ANY FAILURES OCCUR
if the answer is "oneOf [YES]"
Source: TSG Resolute (Community Member)