Monthly PM - Xrite Exact
This procedure outlines the monthly preventive maintenance for Xrite eXact units, including cleaning the optics and calibration plaque, verifying calibration accuracy for DEcmc and density, and recording the results.
Monthly PM - Xrite Exact
This procedure outlines the monthly preventive maintenance for Xrite eXact units, including cleaning the optics and calibration plaque, verifying calibration accuracy for DEcmc and density, and recording the results.
Cleaning the Optics
Cleaning Calibration Plaque
Verify Accurate Calibration (DE2000) - Black
BLACK - Stored
Measure DE2000 against Autura standard: TREF Black
Is DEcmc below 0.6?
If DEcmc is above 0.6, describe recalibration attempts and escalate if necessary.
Verify Accurate Calibration (DE2000) - Cyan
Cyan - Stored
Measure DE2000 against Autura standard: TREF Cyan
Is DEcmc below 0.6?
If DEcmc is above 0.6, describe recalibration attempts and escalate if necessary.
Verify Accurate Calibration (DE2000) - Magenta
Magenta - Stored
Measure DE2000 against Autura standard: TREF Magenta
Is DEcmc below 0.6?
If DEcmc is above 0.6, describe recalibration attempts and escalate if necessary.
Verify Accurate Calibration (DE2000) - Yellow
Yellow - Stored
Measure DE2000 against Autura standard: TREF Yellow
Is DEcmc below 0.6?
If DEcmc is above 0.6, describe recalibration attempts and escalate if necessary.
T-Ref Card Inspection
Does the T-Ref card need to be replaced based on date or cleanlienss?
if the answer is "oneOf [Yes]"
Verify Accurate Calibration (Density) - Black
Does the T-Ref card need to be replaced based on date or cleanlienss?
if the answer is "oneOf [Yes]"
Read BLACK sample of the T-Ref placard and record values.
Is value +/- 0.02 from listed value?
if the answer is "oneOf [No]"
If density reading is above 0.02 from the listed value, describe recalibration attempts and escalate if necessary
Verify Accurate Calibration (Density) - Cyan
Read Cyan sample of the T-Ref placard and record values.
Is value +/- 0.02 from listed value?
if the answer is "oneOf [No]"
If density reading is above 0.02 from the listed value, describe recalibration attempts and escalate if necessary
Verify Accurate Calibration (Density) - Magenta
Read Magenta sample of the T-Ref placard and record values.
Is value +/- 0.02 from listed value?
if the answer is "oneOf [No]"
If density reading is above 0.02 from the listed value, describe recalibration attempts and escalate if necessary
Verify Accurate Calibration (Density) - Yellow
Read Yellow sample of the T-Ref placard and record values.
Is value +/- 0.02 from listed value?
if the answer is "oneOf [No]"
If density reading is above 0.02 from the listed value, describe recalibration attempts and escalate if necessary
Completion
Completed by
Click here to sign
Source: CL&D - ProMach (Community Member)