FB25-14 Monthly Calibration
This procedure outlines the monthly calibration steps for the FB25-14 machine. Using a certified voltage meter, the technician verifies that each power supply is within its specified tolerance to ensure safe and accurate machine operation.
FB25-14 Monthly Calibration
Monthly calibration of the FB25-14 ensures the instrument maintains optimal accuracy and dependable performance by preventing measurement drift and identifying potential issues early. This routine calibration supports compliance with quality and safety standards, minimizes downtime and costly mistakes, and helps extend the overall lifespan of the equipment. It ensures precise, reliable operation that keeps workflows efficient, safe, and audit-ready.
Calibration Procedure
40. Measure the voltage at 24.00vdc power supply. (23.8vdc - 24.2vdc)
Meter name
Unit
50. Measure the voltage at 5.00vdc power supply. (4.8vdc - 5.2vdc)
Meter name
Unit
90. Enter the ID# of the voltage meter that was used.
100. Sign Off
Click here to sign