FB25-14 Monthly Calibration

This procedure outlines the monthly calibration steps for the FB25-14 machine. Using a certified voltage meter, the technician verifies that each power supply is within its specified tolerance to ensure safe and accurate machine operation.
TAC Manufacturing
12/10/2025

FB25-14 Monthly Calibration

Monthly calibration of the FB25-14 ensures the instrument maintains optimal accuracy and dependable performance by preventing measurement drift and identifying potential issues early. This routine calibration supports compliance with quality and safety standards, minimizes downtime and costly mistakes, and helps extend the overall lifespan of the equipment. It ensures precise, reliable operation that keeps workflows efficient, safe, and audit-ready.

    Calibration Procedure

    40. Measure the voltage at 24.00vdc power supply. (23.8vdc - 24.2vdc)

    • FB25-1340.PNG
    Meter name
    Unit

    50. Measure the voltage at 5.00vdc power supply. (4.8vdc - 5.2vdc)

    • FB25-1350.PNG
    Meter name
    Unit

    90. Enter the ID# of the voltage meter that was used.

    100. Sign Off

    Click here to sign

Source: TAC Manufacturing (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.