PR 11.3 Calibration Validation Form

This form records caliper validation in accordance with ASME B89.1.14-2018 and ISO 9001:2015 (7.1.5). Perform validation in a controlled environment at 68°F (20°C) using NIST-traceable reference standards. “Measurement Points” are verification checks at different lengths across the caliper’s measuring range (not “points on the tool”). Enter the reference standard value and the caliper reading. The form calculates Error, applies the correct MPE, and returns PASS/FAIL automatically.
TSG Resolute
01/20/2026

PR 11.3 Calibration Validation Form

This form records caliper validation in accordance with ASME B89.1.14-2018 and ISO 9001:2015 (7.1.5). Perform validation in a controlled environment at 68°F (20°C) using NIST-traceable reference standards. “Measurement Points” are verification checks at different lengths across the caliper’s measuring range (not “points on the tool”). Enter the reference standard value and the caliper reading. The form calculates Error, applies the correct MPE, and returns PASS/FAIL automatically.

    INSTRUMENT INFORMATION

    INSTRUMENT ID

    MANUFACTURER

    MODEL

    SERIAL NUMBER

    CALIPER TYPE

    ENVIRONMENT & AS RECEIVED CONDITION

    AS RECEIVED CONDITION

    AS RECEIVED NOTES (REQUIRED IF FAIL)

    AS LEFT CONDITION

    OUTSIDE JAW (EXTERNAL) MEASUREMENTS

    Outside Jaw Check: Measure NIST-traceable gauge blocks (or certified length standards) using the outside jaws. Complete 5 measurement points across the caliper’s range. The form automatically calculates Error, selects MPE, and returns PASS/FAIL.

    Gauge Block Actual (From Lab) - Point 1

    Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions

    OUTSIDE MEASUREMENT #1 DETAILS

    Measured Value - Point 1

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement Point 1 - PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.001 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    OUTSIDE MEASUREMENT #2 DETAILS

    Gauge Block Actual (From Lab) - Point 2

    Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions

    Measured Value - Point 2

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement Point 2 - PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.001 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    OUTSIDE MEASUREMENT #3 DETAILS

    Gauge Block Actual (From Lab) - Point 3

    Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions

    Measured Value - Point 3

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement Point 3 - PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.001 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    OUTSIDE MEASUREMENT #4 DETAILS

    Gauge Block Actual (From Lab) - Point 4

    To Acquire 11" for the Validation Target, use the 10" Gauge Block (TSG-GB003) paired with the 1" Gauge Block (TSG-GB002)

    • BLOCKS-1.jpeg
    • BLOCKS-2.jpeg

    Measured Value - Point 4

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement Point 4 - PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): • Tolerance: ±0.0015 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    INTERNAL MEASUREMENTS

    INSIDE MEASUREMENT #1 DETAILS

    Gauge Block Actual (From Lab) - ID Point 1

    Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions

    Measured Value - ID Point 1

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement ID Point 1 - PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): •Tolerance: ±0.001 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    INSIDE MEASUREMENT #2 DETAILS

    Gauge Block Actual (From Lab) - ID Point 2

    Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions

    Measured Value - ID Point 2

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement ID Point 2 - PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): •Tolerance: ±0.002 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    INSIDE MEASUREMENT #3 DETAILS

    Gauge Block Actual (From Lab) - ID Point 3

    Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions

    Measured Value - ID Point 3

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement ID Point 3- PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): •Tolerance: ±0.002 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    INSIDE MEASUREMENT #4 DETAILS

    Gauge Block Actual (From Lab) - ID Point 4

    Select the Gauge Block Target to be Measured - Use Gauge Block: TSG-GB001 - Targets have been updated here to Reflect the Certified Dimensions

    Measured Value - ID Point 4

    Calculate Measured Error

    Measured Error is calculated by the following formula : = (Gauge Block Actual (From Lab)) - (Measured Value)

    Measurement ID Point 4- PASS/FAIL

    Acceptance Criteria (ASME B89.1.14-2018): Measured error shall not exceed the manufacturer’s Maximum Permissible Error (MPE): •Tolerance: ±0.0025 in

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FAIL]"

    FINAL DETAILS

    DID ANY FAILURES OCCUR

    if the answer is "oneOf [YES]"
Source: TSG Resolute (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.