Verification Record of Absolute Viscometer

Procedure ACMP-OP-007: Equipment and apparatus used in testing materials must be calibrated or verified for accuracy on a regular basis to assure the equipment is giving reliable results.
Ergon - Lubbock 14842
05/23/2022

Verification Record of Absolute Viscometer

Procedure ACMP-OP-007: Equipment and apparatus used in testing materials must be calibrated or verified for accuracy on a regular basis to assure the equipment is giving reliable results.

    Date

    Performed By

    Manufacturer

    Lab ID

    Model

    Thermometer ID

    Viscometer Bath

    Bath filled with clean distilled water or clear oil

    • Yes
    • No
    • N / A

    Maintains temperature of 60 +/- 0.1 degree C

    • Yes
    • No
    • N / A

    Actual Temp

    Thermometer correctly immersed at test temperature

    • Yes
    • No
    • N / A

    Vacuum System

    Maintains vacuum of 300 +/- 0.5 mmHG

    • Yes
    • No
    • N / A

    Actual mmHg

    Holds vacuum when system is evacuated and closed

    • Yes
    • No
    • N / A

    Standardized at least once per year

    • Yes
    • No
    • N / A

    Thermometer

    ASTM 47C or 47F used

    • Yes
    • No
    • N / A

    Other thermometric device with an accuracy of at least 0.02 degree C used

    • Yes
    • No
    • N / A

    Thermometer or other device calibrated at least every six months

    • Yes
    • No
    • N / A

    Viscometers

    Clean and free of deposits or contaminates

    • Yes
    • No
    • N / A

    Immersed so top timing mark at least 20 mm below surface of bath liquid at test temperature

    • Yes
    • No
    • N / A

    Cleaned with suitable solvent and properly dried or cleaned in a glass cleaning oven not set above 500 degree C

    • Yes
    • No
    • N / A

    Verified with proper viscosity standard at least every five years

    • Yes
    • No
    • N / A

    Date of last verification

    Comments

Source: Ergon - Lubbock 14842 (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Global Procedure Library is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.