Back

Vital Signs and Patient Monitor Inspection and Preventative Maintenance

The Vital Signs and Patient Monitor Inspection and Preventative Maintenance (PM) process is crucial in ensuring that medical devices used to monitor patient health are functioning properly and accurately. This process involves routine inspections, calibrations, and preventive measures to minimize equipment failures, enhance patient safety, and maintain compliance with healthcare regulations.
SLBMC
07/02/2025

Vital Signs and Patient Monitor Inspection and Preventative Maintenance

The Vital Signs and Patient Monitor Inspection and Preventative Maintenance (PM) process is crucial in ensuring that medical devices used to monitor patient health are functioning properly and accurately. This process involves routine inspections, calibrations, and preventive measures to minimize equipment failures, enhance patient safety, and maintain compliance with healthcare regulations.
Procedure Max Score: 6

    Inspect the case, display screen, buttons and knob for obvious signs of damage

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FLAG,FAIL]"

    Reason

    Inspect the power cord, mount bracket and module accessories for obvious signs of damage

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FLAG,FAIL]"

    Reason

    Inspect all external connections for loose connectors, bent pins or frayed cables

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FLAG,FAIL]"

    Reason

    Does this unit have an NIBP module?

    • Yes
    • No
    • N / A
    if the answer is "oneOf [Yes]"

    Inspect NIBP cuffs, connectors and hoses for wear

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [Yes]"

    Do values align with patient simulator?

    • Pass
    • Flag
    • Fail

    Does this unit have an SPO2 module?

    • Yes
    • No
    • N / A
    if the answer is "oneOf [Yes]"

    Inpsect SPO2 Sensor and/or trunk cable for signs of wear and damage

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [Yes]"

    Do values align with patient simulator?

    • Pass
    • Flag
    • Fail

    Does this unit have a temperature module?

    • Yes
    • No
    • N / A
    if the answer is "oneOf [Yes]"

    Inspect sensor for signs of wear and damage

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [Yes]"

    Is module operational?

    • Yes
    • No
    • N / A

    Does this unit have an ECG Module?

    • Yes
    • No
    • N / A
    if the answer is "oneOf [Yes]"

    Inspect ECG cables for signs of wear and damage

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [Yes]"

    Do values align with patient simulator?

    • Pass
    • Flag
    • Fail

    Additional notes

Source: SLBMC (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.