FB25-16 Monthly Calibration
This procedure outlines the monthly calibration steps for the FB25-16 machine. Using a certified voltage meter, the technician verifies that each power supply is within its specified tolerance to ensure safe and accurate operation.
FB25-16 Monthly Calibration
Monthly calibration of the FB25-16 ensures consistent accuracy, reliability, and safety across operations by preventing measurement drift and catching early signs of equipment issues. This proactive process supports compliance with quality and audit standards, reduces downtime and costly errors, and extends the lifespan of the instrument. It provides dependable performance and maintains the high level of precision required for efficient, safe, and compliant workflows.
Calibration Procedure
40. Measure the voltage at 24.00vdc power supply. (23.8vdc - 24.2vdc)
Meter name
Unit
50. Measure the voltage at 5.00vdc power supply. (4.8vdc - 5.2vdc)
Meter name
Unit
90. Enter the ID# of the voltage meter that was used.
100. Sign Off
Click here to sign