Belimed MST-V is a cleaning and sterilization system commonly used in healthcare and laboratory environments. Preventative Maintenance (PM) Level 1 is a fundamental set of tasks aimed at ensuring optimal performance, safety, and reliability of the MST-V system. This level of maintenance typically focuses on routine checks and minor adjustments to prevent potential issues and prolong the lifespan of the equipment.
SLBMC
06/13/2025

Belimed MST-V Preventative Maintenance Level 1

Belimed MST-V is a cleaning and sterilization system commonly used in healthcare and laboratory environments. Preventative Maintenance (PM) Level 1 is a fundamental set of tasks aimed at ensuring optimal performance, safety, and reliability of the MST-V system. This level of maintenance typically focuses on routine checks and minor adjustments to prevent potential issues and prolong the lifespan of the equipment.
Procedure Max Score: 20

    Inspect door drive system

    Is chain sufficiently lubricated and free of dirt?

    • Yes
    • No
    • N / A
    if the answer is "oneOf [No]"

    Clean chain and grease

    • Pass
    • Flag
    • Fail

    Check entire piping system for leaks during operation

    Re-seal leakages, replace gaskets or leaky parts

    • Pass
    • Flag
    • Fail

    Check and sterilize ventilation filter

    Dismantle the ventilation filter F305 and inspect it visually for damage and obstructions.

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FLAG,FAIL]"

    Inspect Chamber Filter

    • Pass
    • Flag
    • Fail
    if the answer is "oneOf [FLAG,FAIL]"

    Leak test verification

    Select and start 'Leak Test' program. When the vacuum has been reached, the next phases up to aeration phase can be skipped with the <cycle advance> button.

    At the beginning of the ventilation phase, clock the time taken for the chamber to be aerated from 100mbar to ambient pressure.

    Does ventilation time exceed 3 minutes?

    • Pass
    • Flag
    • Fail
    if the answer is "unchecked "
    if the answer is "checked "

    Is filter in good condition?

    • Pass
    • Flag
    • Fail
    if the answer is "noneOf [PASS]"

    Inspect fan in switch cabinet

    • Pass
    • Flag
    • Fail
    if the answer is "noneOf [PASS]"

    Inspect filter in switch cabinet

    • Pass
    • Flag
    • Fail
    if the answer is "noneOf [PASS]"

    Perform warm-up and leak test

    • Pass
    • Flag
    • Fail

    Perform integrated bowie dick test

    • Pass
    • Flag
    • Fail

    Department Representative Name

    Signature

    Click here to sign
Source: SLBMC (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.