Autoclave Calibration Verification

If any abnormal operation or condition is noted during the inspection procedures that could compromise safe operation of the equipment, the technician shall immediately remove the equipment from service until repairs have been made.
LSG - Global Enterprise (Facilities)
11/14/2025

Autoclave Calibration Verification

Autoclave Calibration Verification guarantees the sterility of products by ensuring the equipment delivers precise temperature and pressure, while also providing documented proof of compliance with industry regulations.

    CERTIFICATION

    Have the thermocouple / data logger and the sterilizer thermocouple been calibrated by an authorized source (e.g. technician / manufacturer) within the preceding 12 months?

    • Yes
    • No
    • N / A

    DATE OF CALIBRATION

    (Attach a copy of the calibration certificate)

    Add or drag pictures

    TEST EQUIPMENT

    PROCEDURE

    Verify that the test equipment accurately measures both freezing and boiling water (32 degrees F and 212 degrees F respectively).

    • Pass
    • Flag
    • Fail

    Insert the tip of the thermocouple probe at least 2” into an uncooked potato. Place the potato in the center of a normal load in the autoclave.

    • Pass
    • Flag
    • Fail

    Record the: AMBIENT CHAMBER TEMPERATURE

    Record the: INTERNAL POTATO TEMPERATURE

    The CYCLE START begins when the chamber reaches the temperature programmed by the facility / manufacturer and may be predetermined.

      CYCLE START TIME

      CYCLE START TEMPERATURE

      CYCLE START PRESSURE

      Continue to observe, and when the temperature probe in the potato indicates a temperature of 212 degrees F (250 degrees F in California) record the:

        TIME

        CHAMBER TEMPERATURE

        CHAMBER PRESSURE

        Allow the autoclave to run a complete cycle and note the:

          TIME

          CHAMBER TEMPERATURE

          CHAMBER TEMPERATURE

          POTATO INTERNAL TEMPERATURE

          FOR A SUCCESSFUL TEST: THE POTATO MUST REACH AN INTERNAL TEMPERATURE OF AT LEAST 212 DEGREES F (250 DEGREES F IN CALIFORNIA) FOR A MINIMUM OF 30 MINUTES

          Picture of Inspection Completion

          Add or drag pictures

          Equipment Repair Checklist

          Technicians are responsible for the following processes before and after any repair. Note: (when possible) remove equipment/ machinery from food processing area/room before repair is made.

          Signature

          Click here to sign
        Source: LSG - Global Enterprise (Facilities) (Community Member)

        Please note that any procedure, checklist, or other document available in the MaintainX Procedure Hub is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.