Standardization Record for Rotational Viscometer
Maintaining a standardization record for a rotational viscometer ensures the accuracy, reliability, and consistency of viscosity measurements. It helps in verifying that the instrument is functioning correctly and within acceptable limits. Regular standardization enhances traceability, supports compliance with quality standards and regulatory requirements, and minimizes measurement errors. Additionally, it provides documented evidence of calibration and performance, which is essential for audits, troubleshooting, and maintaining product quality.
Standardization Record for Rotational Viscometer
Maintaining a standardization record for a rotational viscometer ensures the accuracy, reliability, and consistency of viscosity measurements. It helps in verifying that the instrument is functioning correctly and within acceptable limits. Regular standardization enhances traceability, supports compliance with quality standards and regulatory requirements, and minimizes measurement errors. Additionally, it provides documented evidence of calibration and performance, which is essential for audits, troubleshooting, and maintaining product quality.
Date
Verified By
NIST Thermometer used
Set Temperature 135 degree C
Time @ 0.5 hr
Temperature Setting
NIST Probe
Time @ 1 hr
Temperature Setting
NIST Probe
Time @ 1.5 hr
Temperature Setting
NIST Probe
Set Temperature 163 degree C
Time @ 0.5 hr
Temperature Setting
NIST Probe
Time @ 1 hr
Temperature Setting
NIST Probe
Time @ 1.5 hr
Temperature Setting
NIST Probe
Set Temperature 190 degree C
Time @ 0.5 hr
Temperature Setting
NIST Probe
Time @ 1 hr
Temperature Setting
NIST Probe
Time @ 1.5 hr
Temperature Setting
NIST Probe
Standard Fluid
Type of fluid
Lot #
Expiration Date
Viscosity Standard
After temperatures have been verified, apply temperature correction to unit set point and check unit with a certified viscosity standard.
Unit capable of maintaining test temperatures withing 1.0 degree C
Temperature 135 degree C
Viscosity Standard N450000
Actual Reading (Visc.)
% Difference (+/- 2%)
Maintenance
Clean the exterior of the thermosel unit with citrus cleaner, removing heat barrier cage if necessary
Clean the temp. control box with citrus cleaner
Check spindles for chips or burrs, remove and replace from service if damaged
Action recommended
Comments