Back

Verification Record for Timing Devices

Procedure ACMP-OP-014% Accuracy = [(A-B)/B]*100A = Reading on lab timer, secB = Reading on atomic clock, sec: Equipment and apparatus used in testing materials must be calibrated or verified for accuracy on a regular basis to assure the equipment is giving reliable results.
Ergon - Lubbock 14842
05/23/2022

Verification Record for Timing Devices

Procedure ACMP-OP-014% Accuracy = [(A-B)/B]*100A = Reading on lab timer, secB = Reading on atomic clock, sec: Equipment and apparatus used in testing materials must be calibrated or verified for accuracy on a regular basis to assure the equipment is giving reliable results.

    Verified By

    Date

    Identification No.

    Serial No.

    Verification Equipment Used

    Atomic Clock

    Starting time

    Ending time

    Timer

    Starting time

    Ending time

    Record the difference between the two

    Percent accuracy

    Tolerance to within 0.05% accuracy or better when tested over an interval of not less than 15 minutes

    • Pass
    • Flag
    • Fail

    Action Recommended

Source: Ergon - Lubbock 14842 (Community Member)

Please note that any procedure, checklist, or other document available in the MaintainX Global Procedure Library is provided for general education and information only and does not constitute legal, medical, or financial advice. MaintainX makes the materials available AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any such use, including compliance with applicable law and with meeting any conditions of product warranties.